ERROR DISPLAYS AND TROUBLESHOOTING
[Note:
If the meter still does not perform normally after the above
measures are taken, call
Jenco
representative.
]
Main
Display
Secondary
Display
Possible cause(s)
Corrective
Action(s)
"over" during
measurements 0.0~100.0
℃
●
Sample
Conductivity
value > 200.0mS;
Sample TDS >
200g/L.
●
Conductivity cell
contaminated or
defective.
●
Incorrect K
constant value
input.
●Sample cannot be
tested
.
●
Decontaminate /
clean cell or
replace cell.
●Input correct K
value.
"over" during
calibration
/
●
Incorrect standard
solution.
●
Conductivity cell
contaminated or
defective.
●
Incorrect K
constant value
input.
●Replace standard
solution.
●Decontaminate /
clean cell or
replace cell.
●Input correct K
value.
"over " during
measurements
over
●
Sample
temperature >100
℃
.
●Reduce sample
temperature.
●
Defective
conductivity cell.
●Replace cell.
undr
●
Sample
temperature
<0.0
℃
●Increase sample
temperature.
●
Defective
conductivity cell.
●Replace cell.