18
D. Save, Recall and Delete Data
a.
Saving readings to memory.
1.
In “Measurement” mode, press the “ENTER” key to save data.
The “Save” icon with the corresponding site number will lit up
for a brief moment to indicate a successful data save.
2.
If the “Full” icon is displayed, this means that all 50 data saving
sites are used up. No new data can be saved until existing
saved data are deleted.
b.
Recalling readings from memory.
1.
To recall saved data, press “ENTER” key at the Recall
interface to go into “Recall” mode.
2.
Press the “UP” or “DOWN” keys to select the storage site
number.
3.
Press “MODE” key to exit “Recall” mode.
c. Deleting
data.
1.
Press the “ENTER” key at the Delete interface to go into
“Delete” mode.
2.
Select “Delete All” or “Delete Each” mode by pressing the “UP”
or “DOWN” key.
3.
In the “Delete all” mode, press “ENTER” key to clear all stored
data. Deletion is now complete.
4.
In the “Delete Each” mode, use “UP” and “DOWN” key to
select data to be deleted. Then press “ENTER” key to delete.
Deletion is now complete. The next set of saved data will
automatically move up a slot in the storage site.
5.
Press “MODE” key to exit “Delete” mode.
Summary of Contents for VisionPlus 9020M
Page 1: ...9020M 02270148 9020M...
Page 2: ...1 2 2 2 3 3 A 3 B 4 5 A 5 B 6 C 7 D 7 8 9 10...
Page 3: ...2 9020M 9020M 9V EEPROM LCD 9020M 50 9020M LD 900G 01 9020M 50 60HZ JENCO 9020M 1 2 3 JENCO...
Page 4: ...3 LCD BAT 1 1 2 9V 3 A 1 2 3 4 5 10 8 9 7 6 11 12 14 13 LCD...
Page 7: ...6 1 2 3 4 B 9020M 1 LD 900G 1 2 MODE...
Page 11: ...10 JENCO 18 201601 57619600 57619008 57619240...