JENCO 6250 Operation Manual Download Page 17

 

16 

will          display  ION , anniciator 1 ( on the higher left 
cornor / model 6250) or red color STAND I LED (model 6219) 
lamp will flash and the conc. unit indicator will on. 1.5 Press the   
STAND/STAND I key, the LCD will display 1.00, anniciator 1 
(on the  higher left conor / model  6250)  or STAND I LED 
lamp (model 6219) will on  and  anniciator 2 (on the  higher 
left cornor/ model 6250) or red color STAND II LED (model 
6219) lamp will flash. The instrument is single point standardized 
and assumes a slope of +58 mV per decade change in 
concentration. 

Er5 will be displayed if the input value e/- 750mV     
(approximately) or the display exceeds 120.0 conc. 
Er2 will be display if the STAND II key is pressed before the     
STAND I key. 
 
1.6 For dual point standardization, remove the electrodes from the         
dilute standard, rinse with the concentrate  standard  or  de-     
ionized water and immerse the electrodes in the concentrate     
standard 2. 
1.7 Press  the  STAND II  key, the anniciator 2 (model 6250) 
or STAND II LED    (model 6219) lamp will on. The    STAND II   
key can be pressed as many times as required for a stable reading 
of 10.00 conc. The instrument is dual point standardized and 
ready for measurements. The electrodes should always be rinsed   
with    the sample or de-ionized water before measure-ments. 
 
Er2 will be displayed if electrode slope small than 50% of     
ideal slope, 58 mV. 
Er5 will be displayed if the input value e/- 750 mV     
(approximately) or the display exceeds 120.0 conc. 

 

 
 
 
 
 

Summary of Contents for 6250

Page 1: ...OPERATION MANUAL MODEL 6250 6219 SERIES MICRO COMPUTER BASED PH mV ION TEMP METER JENCOELECTRONICS LTD MANUFACTURER OF PRECISION INSTRUMENTS...

Page 2: ...touch keys with audio feedback buffer recognition electrode slope recognition ATC Automatic Temperature Compensation capabilities temperature measurement to 0 1 degree C visual and audio operational e...

Page 3: ...you must replace the new batteries when OFF is display ie automatic turn OFF to prevent incorrect measurements 2 AOpen battery compartment by pushing closure up This is most easily open by using a coi...

Page 4: ...3 Figure 1 Battery Compartment...

Page 5: ...4...

Page 6: ...5...

Page 7: ...6...

Page 8: ...9 8 conc ION Accuracy 2 LSD Resolution 0 05 conc range 0 00 to 25 00 conc 0 1 conc range 25 00 to 50 0 conc 0 2 conc range 50 0 to 119 8 conc Operating temperature 0 to 50 Power supply Six AAA batteri...

Page 9: ...the instrument in pH or mV mode But not for ion mode 6 CLEAR key The CLEAR key is used to reset the instruments mode and clear the pH ion calibration values The CLEAR key has a delayed reaction of 2 s...

Page 10: ...robe 1 1 Connect the pH electrode and the ATC Temp to the connectors of the instrument Refer to the Figure 2 The ATC annicator model 6250 green color ATC LED lamp model 6219 will be on Press the MODE...

Page 11: ...ffer 7 00pH or buffer 6 86pH rinse with distilled water and immerse in buffer 4 01pH 10 01pH model 6250 model 6219 or buffer 4 00pH 9 18pH model 6250C model 6219C The instrument will display the tempe...

Page 12: ...iator model 6250 or green color STAND LED lamp Model 6219 will start flash Per Figure 3 2 2 Rinse the pH electrode in distilled water and immerse in buffer 7 00 model 6250 model 6219 or 6 86 model 625...

Page 13: ...the temperature of the second buffer as set in 2 5 Refer to Table 1 If the reading still drifts repeat 2 6 until a stable reading is obtained The micro computer inside the instrument will calculate a...

Page 14: ...pH value measured is out of the 2 000 to 16 000 pH range 2 Measurement with manual temperature compensation in the pH mode 2 1 Connect the pH electrode to the connector of the instrument Refer to Fig...

Page 15: ...rement range 5 0 to 125 0 degree C Once the temperature brought within range Er3 will disappear and the correct temperature reading will be displayed It is not necessary to press the CLEAR key Er4 wil...

Page 16: ...s used to handle ion standards and samples should always be cleaned with deioned water Rule B THe standard solutions and samples should always be at the same temperature Rule C A magnetic stirrer shou...

Page 17: ...20 0 conc Er2 will be display if the STAND II key is pressed before the STAND I key 1 6 For dual point standardization remove the electrodes from the dilute standard rinse with the concentrate standar...

Page 18: ...35 4 020 6 842 6 979 9 107 9 928 40 4 030 6 836 6 973 9 074 9 891 45 4 042 6 832 6 969 9 044 9 859 50 4 055 6 831 6 968 9 017 9 831 The actual reading of the instrument can differ from the values show...

Page 19: ...grams for each computer language and amore detailed explanation of the software driver are included in the accompanying disk PREPARING THE METER FOR OPERATIONS VIA THE RS 232C INTERFACE The Meter come...

Page 20: ...250 MODEL 6219 RS232 PARAMETERS BAUD RATE 2400 FIXED PARITY SOFTWARE DRIVER CONTROLLED DATA BITS 8 STOP BIT 1 PROTOCOL SOFTWARE DRIVER CONTROLLED WARRANTY Jenco Instruments Ltd Warrants this product t...

Page 21: ...will be charged on all unauthorized returns NOTE Jenco Instruments Inc reserves the right to make improvements in design construction and appearance of our products without notice JENCO INSTRUMENTS IN...

Reviews: