14
TEMPERATURE MEASUREMENTS
The two model 6250/6219 can be used to measure temperature
independently with ATC/Temp probe with using the pH electrode.
1.1
If the pH electrode is not connected to the instrument, the pH input
must shorted to the reference input by the protect cap.
1.2 Connect the ATC/Temp probe to the input connector. Refer to
Figure 2. The ATC/MANUAL LED lamp will change from red
color to green color and enter into ATC mode.
1.3 Place the temperature sensor in the media to be measured. The
measured temperature is displayed.
Er3 will be displayed if the temperature is out of the temperature
measurement range -5.0 to 125.0 degree C. Once the temperature
brought within range, Er3 will disappear and the correct
temperature reading will be displayed. It is not necessary to press
the CLEAR key.
Er4 will be displayed if the temperature is out of the temperatue
range 0 to 50.0 degree C, Please press MODE key to enter into mV
or Ion mode for the other measurements.
mV MEASUREMENTS
A combination electrode can be used in place of the working electrode and
the reference electrode. The MANUAL indication will be on if the
ATC/Temp probe is not connected to the instrument.
1.1 Connect the working electrode to the input BNC connector and the
reference electrode to the input Pin connector. Refer figure 2.
1.2 Press the MODE key for the instrument to enter into mV mode
and press RESOL. key to select optimum resolution.