JENCO 6179M Operation Manual Download Page 21

20

ERROR DISPLAYS AND TROUBLESHOOTING

G

[Note:

If the meter still does not perform normally after the above

measures are taken, call

JENCO

representative.

]

Main

Display

Possible cause(s)

Corrective Action(s)

"Er1"

1.

“Stand”

 

was

 

pressed

 

before

 

the

 

electrode

 

and

 

ATC/Temp

 

probe

 

settled

 

to

 

within

 

+/-1.00

 

pH

 

of

 

the

 

buffer

 

value.

2. pH electrode offset is

greater/less than +/-1.00 pH.

3. pH electrode is faulty.

1. Press

“Set/Clear”

key. Allow

sufficient time for the electrode

and ATC/Temp probe to

stabilize. Re-press

“Stand”

key

to start the calibration procedure.

2. Replace the buffer and/or the pH

electrode. Press

“Set/Clear”

key

to recalibrate meter.

3. Replace electrode.

"Er2"

1.

“Slope/Enter”

was pressed

before the electrode and

ATC/Temp probe settled to

within 30% of the buffer value.

2. Buffer 1.68, 4.00, 4.01, 9.18,

10.01 and 12.46 pH is not

correct.

3. pH electrode slope is off by

more than 30% of ideal slope.

1. Allow sufficient time for the

electrode and ATC/Temp probe

to stabilize. Re-press

“Slope/Enter”

key to continue

the calibration procedure.

2. Check if the correct buffer is

used.

3. Replace the buffer and/or the pH

electrode. Press

“Set/Clear”

key

to recalibrate meter.

"Er3"

In pH calibration mode,

temperature is out of the 0.0 to

60.0 °C range.

Bring the pH buffer

temperature within range.

“Er4”

In ion calibration mode, the

slope (mV difference between

two consecutive points) is out

of the 15 to 90 mV range.

1. Replace the standard

solution and/or the ISE.

2. Recalibrate meter.

“Er5”

In ion calibration mode, the end

user exited the calibration

mode before completing the

number of pre-selected

calibration points.

Recalibrate meter.

"over"/

“undr”

1. Measured pH is out of the

-2.00 to 16.00 pH range.

2. Measured mV (RmV) is out of

the –1999.9 to 1999.9 mV

range.

3. Measured ion is out of the 0 to

2000 range.

4. Measured temperature is

out of the 0.0 to 100.0 °C

range.

1. Bring sample pH into the correct

measuring range.

2. Bring sample ORP into the

correct measuring range.

3. Bring sample ion into the

correct measuring range.

4. Bring sample temperature into

the correct measuring range.

Summary of Contents for 6179M

Page 1: ...Operation Manual MODEL 6179M Microcomputer Based pH ORP ION Temperature Benchtop Meter 6179M...

Page 2: ...description 5 OPERATIONAL PROCEDURES 7 A Buffer Set Selection 7 B pH Calibration 7 C pH Measurements 11 D Temperature Measurements 12 E mV offset 12 F mV Rmv Measurements 13 G Ion Calibration 14 H Io...

Page 3: ...lly sense the end point and LOCK the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH mV Rmv ion values resulting in prec...

Page 4: ...E BATTERIES To insert the batteries into the meter follow the procedure outlined below 1 Use a Philip screw driver and unlock the battery cover by turning the screw driver in the counter clockwise dir...

Page 5: ...0 12 11 8 1 68 4 01 7 00 10 01 4 00 6 86 9 18 12 46 Figure 3 Active LCD screen 1 WAIT This will be displayed when the unit is still waiting for a stable reading or end point sensing 7 AUTO AUTOLOCK mo...

Page 6: ...he unit enters into the calibration mode 12 SECONDARY DISPLAY For temperature in B Operational keys description Figure 4 NO Key Description 1 Stand Stand Slope Enter These two keys are used for pH ORP...

Page 7: ...and shut off the meter 5 Mea Eff Stand Slope Mode Clear Mea Eff This key is used to bring the unit out of the AUTOLOCK condition when operating in the pH AUTOLOCK mV RmV AUTOLOCK or ion AUTOLOCK mode...

Page 8: ...Clear key for 5 seconds all LCD elements will lit up The meter clears all calibration values stored in internal memory b Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC...

Page 9: ...pensates for the pH electrode slope deviation corresponding to the values of the calibrated buffers After calibration press and hold Mea Eff key for about 5 seconds to display the new electrode effici...

Page 10: ...l memory b Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC Temp connector of the unit c Press Mode key for the LCD to display pH and the AUTOLOCK annunciator is off ATC...

Page 11: ...AUTOLOCK mode a Turn the unit on In pH NON AUTOLOCK mode press Set Clear key for 5 seconds all LCD elements will lit up The meter clears all calibration values stored in internal memory b Connect the...

Page 12: ...Rinse the pH electrode and ATC Temp probe with distilled water and immerse in the sample to be measured Remove any air bubbles trapped around the probe by shaking or stirring the probe d Press the Mea...

Page 13: ...time for the display to stabilize The meter will display the pH value of the sample at the displayed sample temperature 4 Measurement with manual temperature compensation in the pH NON AUTOLOCK mode...

Page 14: ...a Connect the optional combination ORP electrode to the BNC connector of the unit b Press Mode key until mV or RmV icon MAN AUTO icons lit up The WAIT icon begins to flash c Rinse the ORP electrode w...

Page 15: ...n in the ION AUTOLOCK mode a Turn the unit on In ION AUTOLOCK mode press Set Clear key for 5 seconds all LCD elements will lit up The meter clears all calibration values stored in internal memory b Pr...

Page 16: ...points are completed Note The slope mV difference between two consecutive points is 15 to 90 mV 2 Calibration in the ION NON AUTOLOCK mode a Turn the unit on In ION NON AUTOLOCK mode press Set Clear k...

Page 17: ...on is complete i To continue with the third the fourth and the fifth point calibration repeat step h Press Mode key to exit calibration mode when the desired calibration points are completed Note The...

Page 18: ...bon Dioxide CO2 Perchlorate CIO4 Chloride Cl Potassium K Copper Cu 2 Silver Sulfide Ag S 2 Cyanide CN Sodium Na Fluoride F Surfactant X X Fluoroborate BF4 Water Hardness J Save Recall and Delete Data...

Page 19: ...he selection screen Using the Up or Down key to select YES or no for clearing all stored data To clear all data must select YES and press the Slope Enter key nonE will be displayed after data is compl...

Page 20: ...5 9 39 4 01 7 08 10 25 13 21 10 1 67 4 00 6 92 9 33 4 00 7 06 10 18 13 01 15 1 67 4 00 6 90 9 28 4 00 7 03 10 12 12 80 20 1 68 4 00 6 88 9 23 4 00 7 01 10 06 12 64 25 1 68 4 00 6 86 9 18 4 01 7 00 10...

Page 21: ...ode and ATC Temp probe to stabilize Re press Slope Enter key to continue the calibration procedure 2 Check if the correct buffer is used 3 Replace the buffer and or the pH electrode Press Set Clear ke...

Page 22: ...00 0 C pH Buffer Temperature range 0 to 60 0 C pH calibration Up to 5 point mV calibration 1 point Offset 150 mV Ion calibration 2 to 5 consecutive points 0 1 1 0 10 0 100 0 or 1000 Input impedance 3...

Page 23: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...