JENCO 6179M Operation Manual Download Page 16

15

g. Press

“Slope/Enter”

key. The mV value of the standard

solution will appear on the main display. The “Wait” icon will

flash until the meter detects a stable value. When the

“Wait” icon disappears and the ion calibration value on the

secondary screen moves up to the next value, the first point

calibration is complete and the meter is ready for the

second point calibration.

h. Rinse the ISE in distilled water and immerse it in the second

standard solution. Press

“Slope/Enter”

key. The “Wait”

icon will flash until the meter detects a stable reading.

When the “Wait” icon disappears and the ion calibration

value on the secondary screen moves up to the next value,

the second point calibration is complete and the meter is

ready for the third point calibration.

i.

To continue with the third, the fourth and the fifth point

calibration, repeat step

“g”

. Press

“Mode”

key to exit

calibration mode when the desired calibration points are

completed.

[

Note:

The slope (mV difference between two consecutive

points) is 15 to 90 mV.]

2. Calibration in the ION-NON-AUTOLOCK mode.

a. Turn the unit on. In ION-NON-AUTOLOCK mode, press

“Set/Clear”

key for 5 seconds, all LCD elements will lit up.

The meter clears all calibration values stored in internal

memory.

b. Press the

“Set/Clear”

key to enter the calibration point

selection screen. Use

“Up”

or

“Down”

key to select 2 to 5

point calibration.

c.

Press

“Slope/Enter”

key to enter the monovalent

ion/divalent ion selection screen. Using the

“Up”

or

“Down”

key to select. Press

“Slope/Enter”

key to save.

d. Connect the ISE to the BNC connector of the unit.

e. Press

“Mode”

key for the LCD to display “MAN” icon will lit

up. The "CAL" icon will begin to flash.

f.

Rinse the ISE in distilled water and immerse it in the first

standard solution. Press and hold

“Stand”

key for 2

seconds to calibrate. The “CAL” icon will lit up. The main

display will show “- - - -”, the secondary display will show

corresponding ion calibration value (0.10, 1.0, 10, 100).

Using the

“Up”

or

“Down”

key to select the first point

calibration value. Press

“Slope/Enter”

key to save.

Summary of Contents for 6179M

Page 1: ...Operation Manual MODEL 6179M Microcomputer Based pH ORP ION Temperature Benchtop Meter 6179M...

Page 2: ...description 5 OPERATIONAL PROCEDURES 7 A Buffer Set Selection 7 B pH Calibration 7 C pH Measurements 11 D Temperature Measurements 12 E mV offset 12 F mV Rmv Measurements 13 G Ion Calibration 14 H Io...

Page 3: ...lly sense the end point and LOCK the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH mV Rmv ion values resulting in prec...

Page 4: ...E BATTERIES To insert the batteries into the meter follow the procedure outlined below 1 Use a Philip screw driver and unlock the battery cover by turning the screw driver in the counter clockwise dir...

Page 5: ...0 12 11 8 1 68 4 01 7 00 10 01 4 00 6 86 9 18 12 46 Figure 3 Active LCD screen 1 WAIT This will be displayed when the unit is still waiting for a stable reading or end point sensing 7 AUTO AUTOLOCK mo...

Page 6: ...he unit enters into the calibration mode 12 SECONDARY DISPLAY For temperature in B Operational keys description Figure 4 NO Key Description 1 Stand Stand Slope Enter These two keys are used for pH ORP...

Page 7: ...and shut off the meter 5 Mea Eff Stand Slope Mode Clear Mea Eff This key is used to bring the unit out of the AUTOLOCK condition when operating in the pH AUTOLOCK mV RmV AUTOLOCK or ion AUTOLOCK mode...

Page 8: ...Clear key for 5 seconds all LCD elements will lit up The meter clears all calibration values stored in internal memory b Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC...

Page 9: ...pensates for the pH electrode slope deviation corresponding to the values of the calibrated buffers After calibration press and hold Mea Eff key for about 5 seconds to display the new electrode effici...

Page 10: ...l memory b Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC Temp connector of the unit c Press Mode key for the LCD to display pH and the AUTOLOCK annunciator is off ATC...

Page 11: ...AUTOLOCK mode a Turn the unit on In pH NON AUTOLOCK mode press Set Clear key for 5 seconds all LCD elements will lit up The meter clears all calibration values stored in internal memory b Connect the...

Page 12: ...Rinse the pH electrode and ATC Temp probe with distilled water and immerse in the sample to be measured Remove any air bubbles trapped around the probe by shaking or stirring the probe d Press the Mea...

Page 13: ...time for the display to stabilize The meter will display the pH value of the sample at the displayed sample temperature 4 Measurement with manual temperature compensation in the pH NON AUTOLOCK mode...

Page 14: ...a Connect the optional combination ORP electrode to the BNC connector of the unit b Press Mode key until mV or RmV icon MAN AUTO icons lit up The WAIT icon begins to flash c Rinse the ORP electrode w...

Page 15: ...n in the ION AUTOLOCK mode a Turn the unit on In ION AUTOLOCK mode press Set Clear key for 5 seconds all LCD elements will lit up The meter clears all calibration values stored in internal memory b Pr...

Page 16: ...points are completed Note The slope mV difference between two consecutive points is 15 to 90 mV 2 Calibration in the ION NON AUTOLOCK mode a Turn the unit on In ION NON AUTOLOCK mode press Set Clear k...

Page 17: ...on is complete i To continue with the third the fourth and the fifth point calibration repeat step h Press Mode key to exit calibration mode when the desired calibration points are completed Note The...

Page 18: ...bon Dioxide CO2 Perchlorate CIO4 Chloride Cl Potassium K Copper Cu 2 Silver Sulfide Ag S 2 Cyanide CN Sodium Na Fluoride F Surfactant X X Fluoroborate BF4 Water Hardness J Save Recall and Delete Data...

Page 19: ...he selection screen Using the Up or Down key to select YES or no for clearing all stored data To clear all data must select YES and press the Slope Enter key nonE will be displayed after data is compl...

Page 20: ...5 9 39 4 01 7 08 10 25 13 21 10 1 67 4 00 6 92 9 33 4 00 7 06 10 18 13 01 15 1 67 4 00 6 90 9 28 4 00 7 03 10 12 12 80 20 1 68 4 00 6 88 9 23 4 00 7 01 10 06 12 64 25 1 68 4 00 6 86 9 18 4 01 7 00 10...

Page 21: ...ode and ATC Temp probe to stabilize Re press Slope Enter key to continue the calibration procedure 2 Check if the correct buffer is used 3 Replace the buffer and or the pH electrode Press Set Clear ke...

Page 22: ...00 0 C pH Buffer Temperature range 0 to 60 0 C pH calibration Up to 5 point mV calibration 1 point Offset 150 mV Ion calibration 2 to 5 consecutive points 0 1 1 0 10 0 100 0 or 1000 Input impedance 3...

Page 23: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...

Reviews: