JENCO 3177MB Operation Manual Download Page 9

8

4. Rinse the conductivity probe in distilled water and immerse

them in the standard solution. Allow temperature reading to

stabilize, then press

“Cal”

key to enter the calibration mode.

“CAL” and “ATC” icons appears on the LCD.

[Note:

Press

“Enter”

key to accept any values changes in

each section and automatically advance to the next section. If

there are no changes, the unit accepts the current value and

proceeds to the next section.

]

TDS Constant

TDS is determined by multiplying conductivity (mS/cm) by a

TDS factor. The default factor value is 0.65. To change the

TDS factor, use the

“Up”

or

“Down”

keys to adjust the value

between 0.30 to 1.00. Press

“Enter”

key to save the new

value and go to the next calibration parameter.

Temperature Coefficient

The unit uses the temperature coefficient to calculate

temperature compensated conductivity. The default value is

1.91%. To change the Temperature Coefficient, use the

“Up”

or

“Down”

keys to adjust the value between 0.00 to 4.00%.

Press

“Enter”

key to save the new value and go to the

next calibration parameter.

Temperature Reference

The unit uses the temperature reference value to calculate

temperature compensated conductivity. The default value is

25

. To change the temperature reference, press the

“Up”

or

“Down”

keys to adjust the value between 15 to 25

.

Press

“Enter”

key to save the new value and go to the

next calibration parameter.

Probe Basic Cell Constant

The main display shows the Cell Constant of the conductivity

probe (calibrated previously or default, the deviation range is

70% to 130%). The secondary display shows the current

selected cell constant which is either 0.10 or 0.475. Press the

“Up”

or

“Down”

keys to select the one you want to use. Press

“Enter”

key to save the new value and go to the next

calibration parameter.

Standard Solutions Calibration

Immerse the probe in a standard of known conductivity

solution (see section

Preparing Standard Solutions

),

Summary of Contents for 3177MB

Page 1: ...Operation Manual MODEL 3177MB Cond TDS Salt Temperature Based Bluetooth Benchtop Meter 3177MB...

Page 2: ...splay 4 B Operational Keys Description 5 MODES OF THE METER 6 OPERATIONAL PROCEDURES 7 A Calibration 7 B Conductivity Salinity TDS Measurements 9 C Save Recall And Delete Data 10 D Preparing Conductiv...

Page 3: ...icators The unit prompts the user through calibration and measurement procedures The unit is also equipped with a non volatile memory allowing the user to store 750 different sets of readings This uni...

Page 4: ...fter unlocking the screw take off the battery cover Fig 1 2 Replace the old batteries with new ones and install them on the correct polarity position 3 Put the battery cover back on the meter Use a Ph...

Page 5: ...ndicates Salinity measurement 3 uS cm Microsiemens indicates Conductivity measurement 9 CELL K Indicates conductivity cell constant value 4 CAL This will be displayed when the unit enters into the cal...

Page 6: ...ternal memory In the Recall and Delete modes press Mode Clear key to exit Recall and Delete modes respectively 3 Power BLE Bluetooth Press and hold this key for 2 seconds to power on and shut off the...

Page 7: ...ES and no for data all delete In the Delete One mode view to be deleted data and data site numbers by pressing these keys MODES OF THE METER This instrument has six modes as follow 1 Conductivity mode...

Page 8: ...rature Coefficient Temperature Reference Probe Basic Cell Constant Standard Solutions Calibration and K Value Input To access these sections 1 Turn the unit on In the conductivity TDS salinity mode pr...

Page 9: ...e is 1 91 To change the Temperature Coefficient use the Up or Down keys to adjust the value between 0 00 to 4 00 Press Enter key to save the new value and go to the next calibration parameter Temperat...

Page 10: ...tant to a known value by pressing the Up or Down keys You can now input the K value from 70 to 130 of the probe basic cell constant Press the Enter key to exit calibration and return to normal operati...

Page 11: ...a a Press Mode Clear key until the dEL onE or dEL ALL shows Then press the Enter key to go into the corresponding Delete one or Delete ALL mode b In the Delete ALL mode press the Enter and enters the...

Page 12: ...prepared APP NAVIGATION A App Functions The 3177MB is a conductivity salinity TDS temperature Bluetooth benchtop meter with iOS and Android apps for tablets Functions of the App includes graphing gui...

Page 13: ...7MB is connected to the tablet the device Bluetooth light will blink slowly After the App has been installed calibrate the 3177MB to ensure the best reading quality D Measurement Screen The Measure Sc...

Page 14: ...ata From this screen tap the Calibrate button and follow the step by step instructions to calibrate the 3177MB Fig 2 G Settings Screen The Settings Screen displays the various settings that affects th...

Page 15: ...rmation c From the saved reading detail screen tap the Notes field to add notes d Tap the button to save and exit the detail screen 6 Share and print saved readings a Tap the Saved tab in the navigati...

Page 16: ...d Wait for the App to discover the 3177MB e Tap the Connect button I Notes 1 One meter can be paired to only one App at a time To put an already paired meter into pairing mode one must first close the...

Page 17: ...olled environment This device and its antenna must not be co located or operating in conjunction with any other antenna or transmitter To comply with FCC RF exposure compliance requirements this grant...

Page 18: ...t K constant value input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 10 0 to 120 0 1 Incorrect standard solution 2 Conductivity...

Page 19: ...80 0 ppt 0 1 ppt 0 5 F S Temperature 10 0 to 120 0 C 0 1 C 0 2 C Conductivity calibration 1 point Reference Temperature 15 0 to 25 0 C default at 25 0 C Temperature Coefficient 0 0 to 4 0 default at...

Page 20: ...odel and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss...

Reviews: