LEAK TESTIN
G
St
ep 1
PRE-CLEA
N
IN
G
St
ep 2
1. W
ipe Insertion T
ube
and distal tip immediately after removal from patient using alcohol wipe.
2. Remove
All Parts
Fiberoptic cable Fiberoptic adaptors Video equipment
3. Perform Leak T
est (dry)
Attach leak tester to scope
Pump up to 160mmHg
W
ait one (1) minute
If pressure falls more than 10mmHg in one (1) minute scope fails.
Do not use (contact JEDMED)
If scope passes go to Step 2
Clean scope immediately after use.*
Do NOT allow debris to dry on scope.
Remove all Loose Material Deposits
Hold scope under lukewarm, running tap water
.
Gently wipe with cloth from scope body to distal tip.
Concentrate on insertion tube and distal tip windows.
Do not twist insertion tube!
Use soft toothbrush or similar
on difficult areas.
Rinse thoroughly
.
Pat dry** with cloth.
Immediately go to Step 3
CLEAN
IN
G
St
ep 3
1. Prepare Enzymatic
Detergent / Clean Scope
Follow manufacturer
’s
instructions carefully
.
Do not over-dilute or under-dilute.
Completely immerse scope in detergent bath.
Brush and wipe scope until no visible debris remain.
Continue brushing/wiping for two (2) additional minutes.
Pay special attention to insertion tube, distal tip windows, heavily soiled areas and physician contact areas.
2. Rinse Thoroughly
Immerse scope in clean basin of fresh, lukewarm tap water
.
Agitate gently
. Remove and
rinse under lukewarm running tap water
. Rinse until all
detergent has been removed.
3. Dry Thoroughly
Dry with medical-grade forced air until completely dry**.
4. Re-Inspect
Visibly re-inspect and re-clean if any debris is found.
DISIN
FECTION
(High-Level)
St
ep 4
1. Fully Immerse Scope
in high-level disinfectant solution.
Follow manufacturer
’s
instructions carefully
.
Allow scope to soak specified
time (use timer).
Remove scope from disinfectant.
2. Rinse Thoroughly
Immerse in basin of sterile or clean tap water
.
Agitate gently
.
Discard water
. Repeat above
two (2) additional times.
Rinse
until all disinfectant has been removed.
3. Dry Thoroughly
Dry with medical-grade forced air
until completely dry**.
4. Functionally T
est Scope
ENZOL
®
Enzymatic Detergent
Do N
OT use ult
rasonic cleaner!
STORAGE
St
ep 5
1.
Assure Scope is
Completely Dry
2. Store Scope
in a covered drawer
,
enclosed storage cabinet or original storage case.
Area
should be well-ventilated and dust-free.
Keep insertion tube and angulation section straight.
Cleaning / Disinfecting Procedure
Ergo-Flex Laryngo-Nasopharyngoscopes
* Note: Clean scope immediately after use. Do NOT
allow bio-materials to dry on scope, as cleaning is much more
difficult. Residual bio-burden significantly decreases ef
fectiveness of disinfectants.
Ef
fective cleaning is the most important part of the disinfection process.
** Note: Drying is critically important to minimize formation o
f biofilm. Biofilm will form on any surface containing non-steri
le water
. Biofilms significantly decrease ef
fectiveness of disinfectants.
ENZOL
®
, CIDEX
®
and CIDEX
®
OP
A are registered trademarks of
ADV
ANCED STERILIZA
TION PRODUCTS, Division of Ethicon Inc., a Johnson & Johnson Company
.
Steris 1E
®
is a registered trademark of STERIS Corporation.
www
.JEDMED.com 314-845-3770
Steris 1E
®
Processing
1. Disinfect / Rinse Scope
Follow manufacturer
’s
instructions carefully
.
2. Dry** Thoroughly
Dry with medical-grade forced air until completely dry**.
3. Functionally T
est Scope
All
cleaning
and
disinfecting
to
be
performed only by appropriately-trained individuals.
Always use “Universal Precautions” and protective attire when working with potentially infectious material.
Only the aforementioned methods and products have been validated. Do not deviate.
All
cloths
and
brushes
should
be
clean,
lint-free and single-use.
Be particularly careful not to damage
fiber bundle or optics while cleaning.
Refer to Ergo-Flex Instruction Manual for additional information.
Ef
fective cleaning and disinfecting is
absolutely required to avoid cross- contamination between patients.
12-2015 Rev 2 Copyright © 2013 JEDMED Instrument Company
All rights reserved.
Read and follow these instructions completely!
CIDEX
®
2.4% CIDEX
®
O
PA
Activated Dialdehyde
STERIS
®
Revital-Ox
TM
Resert
®
High-Level Disinfectant
(USA-Australia-New Zealand-
Europe-Middle East)