Jedmed EF-N Instructions For Use Manual Download Page 17

 

LEAK TESTIN

G

St

ep 1

PRE-CLEA

N

IN

G

St

ep 2

1.  W

ipe Insertion T

ube

and distal tip immediately  after removal from patient using alcohol wipe.

2.  Remove 

All Parts

Fiberoptic cable Fiberoptic adaptors Video equipment

3.  Perform Leak T

est (dry)

 Attach leak tester to scope

 Pump up to 160mmHg

 W

ait one (1) minute

If pressure falls more than  10mmHg in one (1) minute scope fails.  

Do not use (contact JEDMED)

If scope passes go to Step 2

   

Clean scope immediately after use.*

Do NOT allow debris to dry on scope.

      Remove all Loose        Material Deposits

Hold scope under lukewarm,  running tap water

.

Gently wipe with cloth from  scope body to distal tip.

Concentrate on insertion  tube and distal tip windows.

Do not twist insertion tube!

Use soft toothbrush or similar 

on difficult areas.

Rinse thoroughly

.

Pat dry** with cloth.

Immediately go to Step 3

CLEAN

IN

G

St

ep 3

1.  Prepare Enzymatic 

 

 

     Detergent / Clean Scope

Follow manufacturer

’s 

instructions carefully

.

Do not over-dilute or under-dilute.

Completely immerse scope in  detergent bath.

Brush and wipe scope until no  visible debris remain.

Continue brushing/wiping for two  (2) additional minutes.

Pay special attention to insertion  tube, distal tip windows, heavily  soiled areas and physician contact  areas.

2.  Rinse Thoroughly

Immerse scope in clean basin  of fresh, lukewarm tap water

.  

Agitate gently

.  Remove and 

rinse under lukewarm running tap water

.  Rinse until all 

detergent has been removed.

3.  Dry Thoroughly

Dry with medical-grade forced air  until completely dry**.

4.  Re-Inspect

Visibly re-inspect and re-clean if  any debris is found.

DISIN

FECTION

(High-Level)

St

ep 4

1.  Fully Immerse Scope 

      in high-level disinfectant solution.

Follow manufacturer

’s 

instructions carefully

.

Allow scope to soak specified 

time (use timer).

Remove scope from disinfectant.

2.  Rinse Thoroughly

Immerse in basin of sterile or  clean tap water

.   

Agitate gently

.  

Discard water

.  Repeat above 

two (2) additional times.  

Rinse 

until all disinfectant has been  removed.

3.  Dry Thoroughly

Dry with medical-grade forced air 

until completely dry**.

4.  Functionally T

est Scope

ENZOL

® 

Enzymatic Detergent

Do N

OT use ult

rasonic cleaner!

STORAGE

St

ep 5

1.  

Assure Scope is 

     Completely Dry

2.  Store Scope

in a covered drawer

enclosed storage cabinet or  original storage case.  

Area 

should be well-ventilated and  dust-free.

Keep insertion tube and  angulation section straight.

Cleaning / Disinfecting Procedure

Ergo-Flex Laryngo-Nasopharyngoscopes

* Note: Clean scope immediately after use.  Do NOT

 allow bio-materials to dry on scope, as cleaning is much more 

difficult.  Residual bio-burden significantly decreases ef

fectiveness of disinfectants.  

              Ef

fective cleaning is the most important part of the disinfection process.

** Note: Drying is critically important to minimize formation o

f biofilm.  Biofilm will form on any surface containing non-steri

le water

.  Biofilms significantly decrease ef

fectiveness of disinfectants.

ENZOL

®

, CIDEX

® 

and CIDEX

®

 OP

A are registered trademarks of 

ADV

ANCED STERILIZA

TION PRODUCTS, Division of Ethicon Inc., a Johnson & Johnson Company

.

Steris 1E

®

 is a registered trademark of STERIS Corporation.

www

.JEDMED.com 314-845-3770

Steris 1E

®

 Processing

1.  Disinfect / Rinse Scope

Follow manufacturer

’s 

instructions carefully

.

2.  Dry** Thoroughly

Dry with medical-grade forced air  until completely dry**.

3.  Functionally T

est Scope

All 

cleaning 

and 

disinfecting 

to 

be 

performed only by appropriately-trained  individuals.

Always use “Universal Precautions”  and protective attire when working with  potentially infectious material.

Only the aforementioned  methods and  products have been validated.  Do not  deviate.

All 

cloths 

and 

brushes 

should 

be 

clean, 

lint-free and single-use.

Be particularly careful not to damage 

fiber bundle or optics while cleaning.

Refer to Ergo-Flex Instruction Manual  for additional information.

Ef

fective cleaning and disinfecting is 

absolutely required to avoid cross- contamination between patients.

12-2015  Rev 2                                                   Copyright © 2013    JEDMED Instrument Company    

All rights reserved. 

       

Read and follow these instructions completely!

CIDEX

®

 2.4%         CIDEX

®

  O

PA

 

Activated Dialdehyde

STERIS

®

 Revital-Ox

TM

 Resert

®

  

High-Level Disinfectant

(USA-Australia-New Zealand-

Europe-Middle East)

Summary of Contents for EF-N

Page 1: ...ervice Manual An Employee Owned Company IMPORTANT Read all instructions before assembling or using this device Instructions for Use Laryngo Nasopharyngoscope 49 4930 EF N 3 4mm 49 4928 EF N Slim 2 8mm...

Page 2: ...s quality system has placed JEDMED as an industry leader As always our focus remains on you our valued customer Building rewarding and lasting relationships is the cornerstone of JEDMED s business phi...

Page 3: ...ion 9 Light Source Risks 10 List of Warnings 11 Functional Test Before Using 12 Ventilation Cap Leak Testing 14 Cleaning Disinfection 17 Maintenance Storage Transport 19 Reorder Information 20 Specifi...

Page 4: ...hen using fiberoptic light sources keep flammables away from light port fiberoptic cable ends and endoscope tip Disconnect from power before cleaning servicing or storing All cleaning and disinfecting...

Page 5: ...e designed for examining the larynx nasal cavity and nasopharynx These areas encompass the upper respiratory tract from the nasal passages to the vocal chords Ergo Flex scopes are intended to be used...

Page 6: ...bundle Handle Leakage tester connector Fiberoptic cable connector Control lever to bend the distal tip Focus ring Eyepiece Insertion tube with bendable distal tip Features Package Contents Scope Leak...

Page 7: ...cable to suitable fiberoptic light source Use only medical grade fiberoptic light sources and light cables to assure patient safety See page 10 11 Warning Karl Storz adaptor removable Wolf adaptor rem...

Page 8: ...Scope MUST be cleaned and disinfected before initial and every use WARNING Users wearing glasses or contact lenses corrected vision Users who remove glasses or contact lenses uncorrected vision Rotat...

Page 9: ...endocoupler Next connect the endocoupler to the scope at the common B cup eyepiece interface Rotate focus ring to align its dot to dot on scope body Focus as needed using focus ring on endocoupler Co...

Page 10: ...ith the scope to determine that the tip temperature remains as cool as possible ideally under 41 C 106 F and that the emitted light does not cause excessive heating of tissue Tip temperatures exceedin...

Page 11: ...fiberoptic light sources video cameras or other equipment attached to the scope MUST be medical grade and type BF floating or electrically isolated connections In addition ALWAYS use an isolated fibe...

Page 12: ...tube and distal tip to verify both are smooth and unblemished 5 Test eyepiece focus adjustment 6 Place scope near representative anatomical target and verify colors observed by naked eye under ambient...

Page 13: ...13 Ergo Flex Operating Service Manual Blank...

Page 14: ...o the inside of the scope When the cap is connected to the scope internal and external atmospheric pressures are allowed to equalize Pressure equalization is required when transporting the scope via a...

Page 15: ...s slightly while inflated 3 Observe gauge If pressure reading drops more than 10 mmHg within 1 minute DO NOT USE scope must be returned to JEDMED for service see page 25 If pressure remains constant a...

Page 16: ...Ergo Flex Operating Service Manual 16 Blank...

Page 17: ...medical grade forced air until completely dry 4 Functionally Test Scope ENZOL Enzymatic Detergent Do NOT use ultrasonic cleaner STORAGE Step 5 1 Assure Scope is Completely Dry 2 Store Scope in a cove...

Page 18: ...Blank...

Page 19: ...ue to reduced atmospheric pressure Refer to page 14 Transport Scope and accessories should be thoroughly cleaned disinfected and dried before storing Keep the instrument lying flat or hanging with the...

Page 20: ...for replacement parts NOTE Replacement Accessory Items Order Description 49 4980 Leakage Tester 49 4985 Ventilation Cap 49 4172 Karl Storz Adaptor 49 4170 Wolf Adaptor 49 4104 ACMI OLYMPUS Adaptor fo...

Page 21: ...king Length 320mm Bend Angle up down 130 Bending Radius 8mm Optical Field of View 80 Depth of Field 1 50mm Weight 230g 8 oz EF N Slim Sheath Diameter 2 8mm Working Length 320mm Bend Angle up down 130...

Page 22: ...Ethicon Inc a Johnson Johnson Company Steris is a registered trademark of STERIS Corporation Product Trade Name High Level Disinfectant Detergent CIDEX OPA CIDEX Activated Dialdehyde Solution 2 4 glu...

Page 23: ...SERVICE INFORMATION...

Page 24: ...parts l Before returning to JEDMED for service instrument must be cleaned disinfected and labelled accordingly If scope has failed leak testing and cannot be immersed for cleaning see next page for re...

Page 25: ...he problem and contact information Return to JEDMED for service A protective plastic sheet is supplied for return of soiled flexible endoscopes 1 Returning Endoscopes for Service An Employee Owned Com...

Page 26: ...Blank...

Page 27: ...t fully inserted into light source or scope connection Broken fibers in fiberoptic cable Debris or fingerprints on eyepiece Clean tip of scope Verify good connections Replace fiberoptic cable Clean ey...

Page 28: ...NOTES...

Page 29: ...5416 JEDMED Court St Louis MO 63129 314 845 3770 phone 314 845 3771 fax e mail info jedmed com website www JEDMED com...

Reviews: