DEVICE OPERATION - List of Warnings

WARNING - 

!

Endoscope tip may become hot depending on the particular fiberoptic light source used.

ALWAYS test tip temperature and heat emitted from scope tip before use.  ALWAYS ensure

either condition does not cause an unacceptable risk to the patient or user.

WARNING - 

!

High-intensity, radiated light may be transmitted from the light emission window at the front of

the scope.  To minimize potential heat-related issues, ALWAYS observe the following:

WARNING - 

!

In the event of a scope malfunction (example: loss of light or angulation function) ALWAYS terminate 

examination as soon as possible and carefully remove scope from patient so as to minimize any risk 

or harm to the patient.

•   Turn off or minimize light source intensity when scope is not in use. 
•   Before use, always test temperature of light emitted from scope tip 

 

to determine adequacy.

•   Do not allow light emitted from scope tip near sensitive tissue or
 

near flammable objects.

•   Do not use more light than required to complete the task at hand.
•   Do not leave light source operating while unattended.

WARNING - 

!

ALWAYS verify any fiberoptic light source used contains the manufacturer’s specified lamp and any 

associated optical filters.

WARNING - 

!

This endoscope is intended to be “floating” or electrically isolated from any other equipment it is attached 

to.  In order to achieve this, any fiberoptic light sources, video cameras or other equipment attached 

to the scope MUST be medical-grade and type BF (floating or electrically isolated connections).  In 

addition, ALWAYS use an isolated fiberoptic cable (never use a fiberoptic cable which contains an 

internal, metal monocoil sheath or other means to electrically connect the two ends of the cable).

WARNING - 

!

Before each use, ALWAYS inspect outer surface of scope to ensure there are no unintended rough 

surfaces, sharp edges or protrusions which could cause harm.

WARNING - 

!

When using a video camera on this endoscope, ALWAYS verify a live video image is being observed (as 

opposed to a stored image in memory) before beginning an examination.  Likewise, after storing a video 

image in memory, always verify a live video image is being observed before continuing examination.

WARNING - 

!

NEVER use this endoscope (or any high-frequency cautery or endotherapy device) in the presence 

of flammable or explosive concentrations of medical gases or anesthetics.

WARNING - 

!

All electrical equipment connected to this equipment will produce patient leakage currents which are 

additive.

WARNING - 

!

In the case where this endoscope is used along with LASER equipment, ALWAYS observe all pertinent 

safety precautions governing laser devices including, but not limited to:

•   User should wear suitable, laser-filtering, eye protection.
•   Never shine a laser beam directly into any person’s eye(s)

WARNING - 

!

Before using with accessories or other endoscopic devices, check compatibility of this scope with the 

criteria for safe use as defined in the equipment’s instructions for use.

Summary of Contents for EF-N

Page 1: ...ervice Manual An Employee Owned Company IMPORTANT Read all instructions before assembling or using this device Instructions for Use Laryngo Nasopharyngoscope 49 4930 EF N 3 4mm 49 4928 EF N Slim 2 8mm...

Page 2: ...s quality system has placed JEDMED as an industry leader As always our focus remains on you our valued customer Building rewarding and lasting relationships is the cornerstone of JEDMED s business phi...

Page 3: ...ion 9 Light Source Risks 10 List of Warnings 11 Functional Test Before Using 12 Ventilation Cap Leak Testing 14 Cleaning Disinfection 17 Maintenance Storage Transport 19 Reorder Information 20 Specifi...

Page 4: ...hen using fiberoptic light sources keep flammables away from light port fiberoptic cable ends and endoscope tip Disconnect from power before cleaning servicing or storing All cleaning and disinfecting...

Page 5: ...e designed for examining the larynx nasal cavity and nasopharynx These areas encompass the upper respiratory tract from the nasal passages to the vocal chords Ergo Flex scopes are intended to be used...

Page 6: ...bundle Handle Leakage tester connector Fiberoptic cable connector Control lever to bend the distal tip Focus ring Eyepiece Insertion tube with bendable distal tip Features Package Contents Scope Leak...

Page 7: ...cable to suitable fiberoptic light source Use only medical grade fiberoptic light sources and light cables to assure patient safety See page 10 11 Warning Karl Storz adaptor removable Wolf adaptor rem...

Page 8: ...Scope MUST be cleaned and disinfected before initial and every use WARNING Users wearing glasses or contact lenses corrected vision Users who remove glasses or contact lenses uncorrected vision Rotat...

Page 9: ...endocoupler Next connect the endocoupler to the scope at the common B cup eyepiece interface Rotate focus ring to align its dot to dot on scope body Focus as needed using focus ring on endocoupler Co...

Page 10: ...ith the scope to determine that the tip temperature remains as cool as possible ideally under 41 C 106 F and that the emitted light does not cause excessive heating of tissue Tip temperatures exceedin...

Page 11: ...fiberoptic light sources video cameras or other equipment attached to the scope MUST be medical grade and type BF floating or electrically isolated connections In addition ALWAYS use an isolated fibe...

Page 12: ...tube and distal tip to verify both are smooth and unblemished 5 Test eyepiece focus adjustment 6 Place scope near representative anatomical target and verify colors observed by naked eye under ambient...

Page 13: ...13 Ergo Flex Operating Service Manual Blank...

Page 14: ...o the inside of the scope When the cap is connected to the scope internal and external atmospheric pressures are allowed to equalize Pressure equalization is required when transporting the scope via a...

Page 15: ...s slightly while inflated 3 Observe gauge If pressure reading drops more than 10 mmHg within 1 minute DO NOT USE scope must be returned to JEDMED for service see page 25 If pressure remains constant a...

Page 16: ...Ergo Flex Operating Service Manual 16 Blank...

Page 17: ...medical grade forced air until completely dry 4 Functionally Test Scope ENZOL Enzymatic Detergent Do NOT use ultrasonic cleaner STORAGE Step 5 1 Assure Scope is Completely Dry 2 Store Scope in a cove...

Page 18: ...Blank...

Page 19: ...ue to reduced atmospheric pressure Refer to page 14 Transport Scope and accessories should be thoroughly cleaned disinfected and dried before storing Keep the instrument lying flat or hanging with the...

Page 20: ...for replacement parts NOTE Replacement Accessory Items Order Description 49 4980 Leakage Tester 49 4985 Ventilation Cap 49 4172 Karl Storz Adaptor 49 4170 Wolf Adaptor 49 4104 ACMI OLYMPUS Adaptor fo...

Page 21: ...king Length 320mm Bend Angle up down 130 Bending Radius 8mm Optical Field of View 80 Depth of Field 1 50mm Weight 230g 8 oz EF N Slim Sheath Diameter 2 8mm Working Length 320mm Bend Angle up down 130...

Page 22: ...Ethicon Inc a Johnson Johnson Company Steris is a registered trademark of STERIS Corporation Product Trade Name High Level Disinfectant Detergent CIDEX OPA CIDEX Activated Dialdehyde Solution 2 4 glu...

Page 23: ...SERVICE INFORMATION...

Page 24: ...parts l Before returning to JEDMED for service instrument must be cleaned disinfected and labelled accordingly If scope has failed leak testing and cannot be immersed for cleaning see next page for re...

Page 25: ...he problem and contact information Return to JEDMED for service A protective plastic sheet is supplied for return of soiled flexible endoscopes 1 Returning Endoscopes for Service An Employee Owned Com...

Page 26: ...Blank...

Page 27: ...t fully inserted into light source or scope connection Broken fibers in fiberoptic cable Debris or fingerprints on eyepiece Clean tip of scope Verify good connections Replace fiberoptic cable Clean ey...

Page 28: ...NOTES...

Page 29: ...5416 JEDMED Court St Louis MO 63129 314 845 3770 phone 314 845 3771 fax e mail info jedmed com website www JEDMED com...

Reviews: