JDS Uniphase T-BERD MTS 8000 Getting Started Manual Download Page 5

Remote Operation Getting Started Guide

Command syntax and format requirements

Transport Module, DMC, MSAM, and 5800 Remote Operation

5

:SENSE:DATA? cstatus:physical:signal

Setup and action values 

are

 case-sensitive.

:SENSe:PAYLoad:BERT:PATTern ALL_ONES

Correct

:SENSe:PAYLoad:BERT:PATTern all_ones

Incorrect

Sockets and

messages

All remote control communications are socket-based TCP/IP. Any message passed via 
remote control must be terminated with a newline character ("\n") as shown (or ENTER in 
direct entry mode).

:INITiate\n

Without the following \n, a message may be interpreted incorrectly or cause an error condi-
tion.

Issuing multiple

commands

You can issue multiple commands on a single line by separating each of the commands 
using a semicolon. The following examples show correct and incorrect presentation of the 
multiple commands.

:INITiate; :SYSTem:ERRor:NEXT?\n

Correct

:INITiate :SYSTem:ERRor:NEXT?\n

Incorrect

Serial command

processing

The remote control system processes commands sequentially in the order they are 
received. It does not process multiple commands simultaneously. If a command is sent 
while another command is still being processed, the second command will not be 
processed until the first command has been fully executed.

Short-form v. Long-

form commands

Commands may be abbreviated using the short form (the capital letters in the command). 
Each word in the command may be either the long or short-form. 

If long-form is used, the entire word must be specified; "partial" long-form is not recognized.

The following examples show the complete long-form version followed by various, short-
form versions of the same command. The final example shows an example of "partial" long-
form that is not acceptable.

:SENSe:DATA? CSTatus:PHYSical:SIGNal

Correct

:SENSe:DATA? CST:PHYS:SIGN

Correct

:SENS:DATA? CSTatus:PHYS:SIGNal

Correct

:SENS:DATA? CST:PHYS:SIGN

Correct

:SENS:DATA? CST:PHYS:SIGNa

Incorrect

Summary of Contents for T-BERD MTS 8000

Page 1: ...applications Controlled modules may be a part of an MTS 8000 or MTS 6000A Test Platform As these platforms are physically different only those physical features that apply directly to remote operatio...

Page 2: ...modules the Transport Module and the MSAM Assembly Used to refer to a complete set of components assembled as an instru ment and used for testing This manual supports three assemblies The Transport Mo...

Page 3: ...the product name connector panel and the Transport Module and MSAM user interfaces to identify the connectors and applications used for 10 7G 11 05G or 11 1G OTN testing PIM The physical interface mo...

Page 4: ...actor Pluggable optical transceiver A variety of optional XFPs are available for testing 10 Gigabit fibre circuits Remote control flow When running a remote control session the basic flow is as follow...

Page 5: ...STem ERRor NEXT nCorrect INITiate SYSTem ERRor NEXT nIncorrect Serial command processing The remote control system processes commands sequentially in the order they are received It does not process mu...

Page 6: ...g used in the test These are indi cated as side and slice These are required arguments and must be filled with a value as follows side PWRS Power Side specify this location when connecting to an MSAM...

Page 7: ...side slice BERT example MOD FUNC SEL PWRS SLIC1 BERT The module will return OFF or ON 2 If the module state response was OFF turn ON the module MOD FUNC SEL side slice BERT ON 3 Query the module s por...

Page 8: ...all currently running applicaitons will be shut down GUI Mode Alternatively remote control may be used without deactivating the GUI This mode can be used to debug remote control tests and to verify th...

Page 9: ...Other common commands from the SCPI standard starting with are not supported REM The REM command is usually the first command issued once connected to the remote control port This puts the unit into r...

Page 10: ...ameter to launch 10 100 1000M Ethernet Layer 2 Traffic Terminate Keep in mind the GUI command may take some time to process The 0 No error message may not show up for a minute or two GUI TermEth10ML2T...

Page 11: ...he table below possibly followed by more information It is recommended that this command is sent automatically after any query command ending with a If a command is errored checking SYSTem ERRor will...

Page 12: ...ions that are available to a specific port The first parameter is reserved for internal use use the value none The second parameter specifies the port to query 1 or 2 Example SYSTem APPLication APPLic...

Page 13: ...cific running instance of an application which directs all subse quent commands to that application It must be sent after the application is launched and the unique id for that application instance ha...

Page 14: ...mand starts the test if it is in the stopped state ABORt The ABORt command stops the test if it is in the running state To perform a test restart and restart all test measurement results perform a tes...

Page 15: ...be space separated from the command String values String values are used for setups where the exact string is not a value defined by the SCPI implementation String values must be enclosed in double qu...

Page 16: ...ons A unique SCPI command exists for every action The examples in this section will show how actions can be implemented and status can be queried and toggled using the same root command with minor var...

Page 17: ...results are queried using the following command SENSe DATA result name where SENSe is an optional entry line represents the port number It is not necessary to enter the port number unless a dual port...

Page 18: ...by semicolons in the order in which they were added to the group Result group examples The following example creates and queries a small group of Ethernet results SENSe DATA GROUP summary CSTatus PHYS...

Page 19: ...t connection If the command returns a value the response is shown on the next line Connecting to an instrument and obtaining the remote control port number on page 19 Starting Remote Control on page 2...

Page 20: ...IC1 BERT Optional query to verify the module is powered on returns OFF or ON ON MOD FUNC PORT BOTH SLIC1 BERT Query for the module s port number 8002 2 Open a TCP IP socket connection to the unit on t...

Page 21: ...o an instru ment and obtaining the remote control port number on page 19 remote control mode may be activated with the REM command It is recommended that the SYSTem ERRor command be sent after every n...

Page 22: ...122 SYSTem APPLication LAUNch Get the identifier of the most recently launched application TermEth1GL3Traffic_122 Launching an application Once the test unit is in remote control mode the following co...

Page 23: ...expected Doing so will provide vital information in the event of a command error or connection issue Restarting the test To restart the test using remote control send the test stop command followed b...

Page 24: ...ure each of those SENSe TEST ENABle OFF Sets the test type to Not Timed SENSe TEST ENABle ON Sets the test type to Timed Test SENSe TEST DURation 3600 Set the timed test duration in seconds SENSe TEST...

Page 25: ...l session to the various platforms They are all similar only differing in the way the script arguments are processed RC Transport pl to connect to a Transport Module in a T BERD MTS 8000 Usage perl RC...

Page 26: ...Optical Innovation Communications Test and Measurement Solutions and the logo are trademarks of JDS Uniphase Corporation Windows is a registered trademark of Microsoft Corporation in the United State...

Reviews: