Remote Operation Getting Started Guide

About the T-BERD/MTS 8000, 6000A, and 5800

2

Transport Module, DMC, MSAM, and 5800 Remote Operation

About the T-BERD/MTS 8000, 6000A, and 5800

Before beginning a remote session, you should be familiar with the products and modules, 
found in the 

T-BERD

MTS 8000 and T-BERD

MTS 6000A Dual Module Carrier, 

Transport Module, and MSAM Getting Started Guide

 and 

T-BERD/MTS 5800 Getting 

Started Manual

.

The following terms have a specific meaning when referring to the T-BERD/MTS 8000 or 
6000A:

10/100/1000 Ethernet

 — Used to represent 10/100/1000 Mbps Ethernet. The 

connector panel also uses 10/100/1000M to identify the connector used for 10/100/
1000 Mbps Ethernet testing.

1GigE

 — Used to represent 1 Gigabit Ethernet. The Transport Module connector 

panel also uses GigE to identify the optical connector used for 1 Gigabit Ethernet 
testing.

10GigE

 — Used to represent 10 Gigabit Ethernet. The Transport Module connector 

panel also uses 

10GE

 and 

10G

 to identify the optical connectors used for 

10 Gigabit Ethernet LAN and WAN testing.

– Action

 — Refers to an item that triggers some change to the test interface, such as 

inserting an alarm or enabling the laser. In the user interface, these are located in the 
Actions toolbar at the bottom of the main results screen.

Application module

 — Used to refer to the component that provides test functionality 

to the assembled instrument. This manual supports two application modules: the 

Transport Module

, and the 

MSAM

.

– Assembly 

— Used to refer to a complete

 set of components

 assembled as an instru-

ment and used for testing. This manual supports three assemblies: The 

Transport 

Module assembly

, consisting of an T-BERD

MTS 8000 base unit and Transport 

Module, the 

MSAM assembly

, consisting of a MSAM, Physical Interface Modules 

(PIMs), and a T-BERD

MTS 6000A base unit, and a 

DMC assembly

, consisting of 

up to two MSAMs, up to four PIMs, a Dual Module Carrier (DMC), and a 
T-BERD

MTS 8000 base unit.

Base unit

 — The unit which connects to the application module and power adapter, 

providing the user interface and a variety of connectivity and work flow tools. If 
optioned to do so, the base unit also allows you to measure emitted power, received 
power, and optical link loss on fiber optic networks.

Battery Module —

 The module connected to the back of the T-BERD / MTS 8000 

base unit, which supplies power whenever it is not provided using the power adapter.

Component

— Used to refer to an individual hardware 

component

 which is connected 

to the other components to build a test instrument (assembly). This manual supports 
the following components: the Transport Module, the MSAM, and the DMC. The base 
units are documented in separate manuals.

– DMC 

— Dual Module Carrier. The DMC is a two slot chassis which you can connect to 

the T-BERD / MTS 8000 base unit to test using up to two MSAM application modules 
and four Physical Interface Modules (PIMs).

FC

 — Used to represent Fibre Channel on the Transport Module connector panel and 

the Transport Module and MSAM user interface to identify the optical connectors used 
for Fibre Channel testing.

Summary of Contents for T-BERD MTS 8000

Page 1: ...applications Controlled modules may be a part of an MTS 8000 or MTS 6000A Test Platform As these platforms are physically different only those physical features that apply directly to remote operatio...

Page 2: ...modules the Transport Module and the MSAM Assembly Used to refer to a complete set of components assembled as an instru ment and used for testing This manual supports three assemblies The Transport Mo...

Page 3: ...the product name connector panel and the Transport Module and MSAM user interfaces to identify the connectors and applications used for 10 7G 11 05G or 11 1G OTN testing PIM The physical interface mo...

Page 4: ...actor Pluggable optical transceiver A variety of optional XFPs are available for testing 10 Gigabit fibre circuits Remote control flow When running a remote control session the basic flow is as follow...

Page 5: ...STem ERRor NEXT nCorrect INITiate SYSTem ERRor NEXT nIncorrect Serial command processing The remote control system processes commands sequentially in the order they are received It does not process mu...

Page 6: ...g used in the test These are indi cated as side and slice These are required arguments and must be filled with a value as follows side PWRS Power Side specify this location when connecting to an MSAM...

Page 7: ...side slice BERT example MOD FUNC SEL PWRS SLIC1 BERT The module will return OFF or ON 2 If the module state response was OFF turn ON the module MOD FUNC SEL side slice BERT ON 3 Query the module s por...

Page 8: ...all currently running applicaitons will be shut down GUI Mode Alternatively remote control may be used without deactivating the GUI This mode can be used to debug remote control tests and to verify th...

Page 9: ...Other common commands from the SCPI standard starting with are not supported REM The REM command is usually the first command issued once connected to the remote control port This puts the unit into r...

Page 10: ...ameter to launch 10 100 1000M Ethernet Layer 2 Traffic Terminate Keep in mind the GUI command may take some time to process The 0 No error message may not show up for a minute or two GUI TermEth10ML2T...

Page 11: ...he table below possibly followed by more information It is recommended that this command is sent automatically after any query command ending with a If a command is errored checking SYSTem ERRor will...

Page 12: ...ions that are available to a specific port The first parameter is reserved for internal use use the value none The second parameter specifies the port to query 1 or 2 Example SYSTem APPLication APPLic...

Page 13: ...cific running instance of an application which directs all subse quent commands to that application It must be sent after the application is launched and the unique id for that application instance ha...

Page 14: ...mand starts the test if it is in the stopped state ABORt The ABORt command stops the test if it is in the running state To perform a test restart and restart all test measurement results perform a tes...

Page 15: ...be space separated from the command String values String values are used for setups where the exact string is not a value defined by the SCPI implementation String values must be enclosed in double qu...

Page 16: ...ons A unique SCPI command exists for every action The examples in this section will show how actions can be implemented and status can be queried and toggled using the same root command with minor var...

Page 17: ...results are queried using the following command SENSe DATA result name where SENSe is an optional entry line represents the port number It is not necessary to enter the port number unless a dual port...

Page 18: ...by semicolons in the order in which they were added to the group Result group examples The following example creates and queries a small group of Ethernet results SENSe DATA GROUP summary CSTatus PHYS...

Page 19: ...t connection If the command returns a value the response is shown on the next line Connecting to an instrument and obtaining the remote control port number on page 19 Starting Remote Control on page 2...

Page 20: ...IC1 BERT Optional query to verify the module is powered on returns OFF or ON ON MOD FUNC PORT BOTH SLIC1 BERT Query for the module s port number 8002 2 Open a TCP IP socket connection to the unit on t...

Page 21: ...o an instru ment and obtaining the remote control port number on page 19 remote control mode may be activated with the REM command It is recommended that the SYSTem ERRor command be sent after every n...

Page 22: ...122 SYSTem APPLication LAUNch Get the identifier of the most recently launched application TermEth1GL3Traffic_122 Launching an application Once the test unit is in remote control mode the following co...

Page 23: ...expected Doing so will provide vital information in the event of a command error or connection issue Restarting the test To restart the test using remote control send the test stop command followed b...

Page 24: ...ure each of those SENSe TEST ENABle OFF Sets the test type to Not Timed SENSe TEST ENABle ON Sets the test type to Timed Test SENSe TEST DURation 3600 Set the timed test duration in seconds SENSe TEST...

Page 25: ...l session to the various platforms They are all similar only differing in the way the script arguments are processed RC Transport pl to connect to a Transport Module in a T BERD MTS 8000 Usage perl RC...

Page 26: ...Optical Innovation Communications Test and Measurement Solutions and the logo are trademarks of JDS Uniphase Corporation Windows is a registered trademark of Microsoft Corporation in the United State...

Reviews: