Troubleshooting 

If any problem described in this section persists, contact JDS Uniphase or your local 
representative. 
 
Every backreflection measurement is affected by the loss before the DUT and by reflections 
before and after the DUT. 
 
 

• 

Loss Before the DUT

 

When a DUT is connected to the Backreflection Meter, the loss from connectors before the DUT 
affects the backreflection reading. Because the light is going out from the Backreflection Meter 
and then returning, it goes through the connectors twice, so the effect of the loss is doubled. If 
the loss is close to 0.15 dB, its effect is compensated automatically by the meter. If the loss 
differs substantially from 0.15 dB, follow the measurement procedures described at the end of 
the 

Measuring Backreflection

 section. 

 

• 

Reflections Before the DUT

 

When a device is connected to the Backreflection Meter, the reflections from the connectors 
affect the backreflection reading. The front panel FC/PC connections must have backreflection 
levels lower than -45 dB for single mode (-25 dB for multi mode), and FC/APC connections must 
have backreflection levels lower than -60 dB for single mode (-40 dB for multi mode). To ensure 
that the Backreflection Meter automatically compensates for this backreflection, follow the setup 
procedure described in the 

Setting Up the Meter for Backreflection Measurements

 section, 

and ensure that the backreflections do not change from the time of the setup. 
 

• 

Reflections After the DUT

 

When making a backreflection measurement, the fiber after the DUT must be terminated to 
eliminate the reflections from the end of the fiber (see the 

Termination Techniques

 section).  

For greatest accuracy when making very low backreflection measurements, terminate near the 
DUT to eliminate reflections from the fiber itself. 
 

• 

Front Panel Connectors

 

Follow the maintenance procedure described in the 

Maintaining the Meter

 section, to ensure 

that the internal pigtail connectors are clean and properly connected to the front panel mating 
sleeve. 
 

• 

Connector Loss and Backreflection

 

Ensure that the loss and backreflection of all the connectors are low and stable, complying with 
the connector specifications. 
 

• 

Internal Backreflection (for single mode model only, except 980 nm model)

 

To confirm proper operation of the Backreflection Meter, terminate the internal cable while using 
the factory-set BR

0

 value (BR

0

 key lamp is off). The backreflection reading must be less than     

-65 dB. 
 

• 

Long Cables (for single mode model only, except 980 nm model)

 

Terminate the cables near the devices or connectors being tested in order to eliminate 
backreflections from the fiber. 

 

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for RM3 Series

Page 1: ...uy your excess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing R...

Page 2: ...RM3 SERIES BACKREFLECTION METER User s Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 3: ...ii 10112341 Rev 004 November 2003 2003 JDS Uniphase All rights reserved Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 4: ...3333 E mail support jdsu com Outside North America Toll Free 800 4069 5599 Toll Free Fax 800 7777 5378 China Toll Free 10 800 140 5599 Toll Free Access Codes by Country Country Code Australia 0011 Eu...

Page 5: ...iv Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 6: ...ty Labeling x General Information and Specifications 1 General Information 1 Backreflection Measurements 2 Loss and Power Measurements 2 Output Port 3 Internal Light Sources 3 Hybrid Jumpers 3 Coheren...

Page 7: ...al 19 Messages and Symbols 20 Calibrating the Meter 21 Backreflection Calibration Adjustment 21 Power Calibration Adjustment 22 Maintaining the Meter 22 Cleaning the Unit 22 Cleaning the Connector End...

Page 8: ...er is 30 VA The unit can also be operated using its internal 6V DC rechargeable battery Fuse Type The fuse type used by the unit is 5 x 20 mm F1A 250 V Safety Instructions The following safety instruc...

Page 9: ...n Table 2 For the exact specification of the lasers described in this manual please refer to the Specifications section Table 2 Laser Class and Parameter Limits Parameter Specification Wavelength 850...

Page 10: ...hazard that can result in serious personal injury Do not operate the unit if an interruption to the protective grounding is suspected In this case ensure that the unit remains inoperative Use only the...

Page 11: ...entification label Figure 1 is placed on the the unit showing name of manufacturer address and month and year of manufacture Figure 1 Identification Label A note as shown in Figure 2 indicating compli...

Page 12: ...xi Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 13: ...xii Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 14: ...3 Series Backreflection Meter An important feature of the Backreflection Meter is the FC APC ultra low backreflection connector at its output port SC APC connector also offered By switching hybrid jum...

Page 15: ...external sources BR is then calculated using the following equation where BR0 is the stored value of the total backreflection up to the device under test DUT and User CAL is the user set calibration...

Page 16: ...jumper for measurement purposes only use the calibrated jumper for calibration purposes only The calibrated jumper has a label attached to it identifying it as the calibrated jumper Coherence Length...

Page 17: ...d when ordered FC PC FC APC or SC PC SC APC hybrid measurement jumper FC PC FC APC or SC PC SC APC hybrid calibrated jumper FC type or SC type detector adapter Detector cap Foot pedal MCTS software an...

Page 18: ...accuracy backreflection power 0 4 dB3 0 05 dB 5 dB loss 0 15 dB 5 dB loss 5 0 7 dB4 0 15 dB 5 dB loss 6 7 Output connector type Panel mounted low backreflection APC Remote interface RS232 or IEEE 488...

Page 19: ...A 250 V Physical Display 16 character LCD Dimensions W x H x D 26 x 11 x 26 cm Weight 4 kg Environmental Operating temperature 0 to 40 C Storage temperature 40 to 70 C Humidity maximum 95 RH from 0 to...

Page 20: ...pment is complete 2 Inspect the unit for structural damage that can have occurred during shipping 3 Connect the unit to the power source using the power cord provided 4 Set the power switch to ON to i...

Page 21: ...do not expose the unit to any source of excessive moisture Cleaning Connectors Caution Connecting damaged or dirty fibers to the unit can damage the connectors on the unit Never force an optical conne...

Page 22: ...off the remaining parts of the connector 6 With the other end of the pipe cleaner or swab dry the areas cleaned 7 Using the dusting gas or compressed air blow the areas cleaned Powering Up the Meter T...

Page 23: ...10 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 24: ...Keys and LEDs Key LED Description ON OFF Power on ON OFF switch BR Sets the meter to Backreflection measurement mode STORE BR0 Stores the BR0 value used in backreflection measurements and toggles the...

Page 25: ...points need to be made as close as possible to the DUT to minimize the errors associated with Rayleigh backscattering in the fiber The backreflection of all components and connections between these tw...

Page 26: ...and occasionally wind the cable in the opposite direction figure 5a Figure 9a Terminating the Jumper for BR Measurement for Single mode except 980nm 2 The second termination is made after the DUT in o...

Page 27: ...e 3 Press the BR key to set the meter to Backreflection mode which is displayed as BR on the meter 4 Clean the output port on the front of the meter and the APC connector green boot of the measurement...

Page 28: ...nt steady The meter displays the backreflection between the two termination points that is caused by the fiber all connections and the DUT This area is shown within the dashed line in Figure 6a for si...

Page 29: ...ange limit last 5 dB of range if using BR0 below 60 dB otherwise to indicate that the setup and measurement procedures described in the previous two sections must be followed carefully to ensure accur...

Page 30: ...D key The key lamp lights to indicate that the user stored value of ID the dark signal from extraneous sources is used to calculate the DUT loss or power If the detector is not sufficiently covered th...

Page 31: ...Power key to set the meter to Relative Power mode In Relative Power mode the meter displays P at the beginning of the reading and rel at the end for example P 0 15 dB rel 2 Press the key to select the...

Page 32: ...is supplied with a foot pedal for convenient data logging to either a computer or printer Data logging operates at a baud rate of 2400 To use the foot pedal 1 Ensure the computer or printer is connect...

Page 33: ...han the minimum value P is greater than the maximum value BRo STORED XX X Displayed momentarily when the BR value XX XdB is being stored as BR0 ID STORED Displayed momentarily when measuring ID TOO MU...

Page 34: ...hich the calibration check is to be performed Choose either wavelength as you will check both 4 Clean the output connector of the calibrated jumper The backreflection value is displayed The value disp...

Page 35: ...n the front panel of the Backreflection Meter The value displayed on the Backreflection Meter must match the value measured by the power meter to within 0 3 dBm If the values do not match to within 0...

Page 36: ...onnector onto the panel 5 Reinstall the connector panel To avoid damaging the input and output port fibers make one or two large loops in the fibers when replacing the panel Recharging the Battery The...

Page 37: ...ollow the setup procedure described in the Setting Up the Meter for Backreflection Measurements section and ensure that the backreflections do not change from the time of the setup Reflections After t...

Page 38: ...a long time Set the power switch to OFF connect the Backreflection Meter to a power source and then set the power switch to ON The BR measurement is approaching the range limit of the Backreflection M...

Page 39: ...26 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 40: ...ght bits per character two stop bits and no parity bit RS232 Pin Assignment The RS232 pin assignment is shown in Figure 15 Figure 15 RS232 Pin Assignment RS232 Interface Specifications The RS232 inter...

Page 41: ...measurement The transmission also includes the wavelength at which the measurement was made and the characters CR LF The character commands that control the meter via the RS232 serial interface port a...

Page 42: ...Left DIP Switch The settings for the left DIP switch are shown in Table 10 Table 10 Left DIP Switch Settings Switch Setting Meaning GPIB ADDR 01000 address 2 CONFIG ON use large serial buffer enable S...

Page 43: ...if the command was accepted 230 PRINT WILL NOT DISPLAY BR 240 STOP 250 FOR I 1 TO 5 Read and display five BR measurements 260 PRINT 1 G 270 INPUT 1 IN 280 PRINT IN 290 NEXT I 300 PRINT 1 A Go to Power...

Page 44: ...ee documents included with this manual RM Series Backreflection and Loss Meter Test Report Certificate of Calibration Hybrid Jumper Backreflection Calibration Artisan Technology Group Quality Instrume...

Page 45: ...32 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 46: ...ained prior to shipping any material to JDS Uniphase The owner s name and address the model number and full serial number of the unit the RMA number and an itemized statement of claimed defects must b...

Page 47: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Reviews: