23
Cleaning the Connector Ends
Clean all connector ends with a lint-free tissue and alcohol before every mating. See the
Cleaning Connectors
section.
1. Loosen the retaining screws of the connector panel, and remove the panel carefully to
access the internal connectors (Figure 14).
Figure 14: Removing the Connector Panel
2. Remove the connector from the mating sleeve in the panel.
3. Clean the connector end faces and mating sleeve in accordance with the
Cleaning
Connectors
section.
4. Reinstall the connector onto the panel.
5. Reinstall the connector panel. To avoid damaging the input and output port fibers, make one
or two large loops in the fibers when replacing the panel.
Recharging the Battery
The letter B is displayed on the Backreflection Meter when the battery power is low. To recharge
the battery:
1. Connect the meter to the AC source.
The Backreflection Meter can be used while it is recharging.
Replacement of Internal Fuses
The following fuses can be replaced by JDS Uniphase qualified technicians only:
1. Fuse F1 on battery protection circuit PCB, 250V, F2A.
2. Fuse on battery cord, 250V, F1.6/10V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for RM3 Series
Page 5: ...iv Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 12: ...xi Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 23: ...10 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 39: ...26 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 45: ...32 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...