Appendix A Test Results
OAM results
HST-3000 Ethernet Testing User’s Guide
281
OAM results
describes the Service Layer OAM CCM results,
such as the number of RDI seconds, loss of continuity indi-
cator, and the number of transmitted and received CCM
frames. Continuity Check must be turned On to observe these
results.
Table 24
S-OAM CCM test results
Result
Description
MD Level
Displays the maintenance domain level
configured for the CCM frame received.
Maint ID
Displays the maintenance association
ID configured for the CCM frame
received.
Peer MEG End
Point ID
Displays the maintenance entity group
end point ID for the instrument’s peer
as configured.
Total Tx Frames
Count of the total number of CCM
frames transmitted since the last OAM
setting was specified or changed.
Total Rx Frames
Count of the number of CCM frames
received since the last OAM setting
was specified or changed.
RDI
Indicates whether or not remote defect
indication is ON or OFF.
RDI Seconds
Count of the number of seconds during
which an RDI was declared since start-
ing or restarting the test.
Loss of Continuity
ON indicates that a loss of continuity
has occurred.
Summary of Contents for IP Video Test Option HST-3000
Page 1: ...HST 3000 Ethernet Testing User s Guide ...
Page 2: ......
Page 22: ...About This Guide Conventions xx HST 3000 Ethernet Testing User s Guide ...
Page 138: ...Chapter 3 Ethernet Testing MAC in MAC testing 116 HST 3000 Ethernet Testing User s Guide ...
Page 202: ...Chapter 5 TCP UDP Testing Loopback testing 180 HST 3000 Ethernet Testing User s Guide ...
Page 278: ...Chapter 9 Troubleshooting Resolving problems 256 HST 3000 Ethernet Testing User s Guide ...
Page 346: ...Glossary 324 HST 3000 Ethernet Testing User s Guide ...
Page 354: ...Index 332 HST 3000 Ethernet Testing User s Guide ...
Page 355: ......