32
V-530 measurement parameters
V-550/560/570
measurement
parameters
Phtmtic mode
Sam
Phtmtrc mode
Sam
Response
Fast
Response
Fast
Band Width
2.0 nm
PMT Voltage
200 V
Note: PMT voltage
varies
with the instrument. Set accordingly.
(5)
Return to the single beam measurement screen and set the wavelength to 500 nm
using the [Goto WL] key on the operation panel.
(6)
Turn the WI lamp stopper adjustment screw (Fig 4.9 or Fig, 4.10) and search for a
position where the data value (value displayed on the CRT) is maximum.
(7)
Replace the inner lid and outer lid of the light source unit.
<<
Adjustment of light source mirror for deuterium lamp
>>
(1)
Turn on the instrument.
(2)
Select [Envir.] from the main menu to display the [Environment] menu.
(3)
Select [Single Beam] from the [Environment] menu to display the single beam
measurement screen. (Refer to Section 9-5 of the Operations manual.)
(4)
Display the parameter screen and set parameters as follows.
V-530 measurement parameters
V-550/560/570
measurement
parameters
Phtmtic mode
Sam
Phtmtrc mode
Sam
Response
Slow
Response
Slow
Band Width
2.0 nm
PMT Voltage
300 V
(5)
Return to the single beam measurement screen and set the wavelength to 250 nm
using the [Goto WL] key on the operation panel.
(6)
Turn the D
2
lamp stopper adjustment screw (Fig 4.9 or Fig, 4.10) and search for a
position where the data value (value displayed on the CRT) is maximum.
(7)
Replace the inner lid and outer lid of the light source unit.