SPECIFICATIONS
Pump OC-15
Specifications
Size
27.5 x 12.0 x 15.0 cm
Weight
2.0 kg
Pressure range
20 ~ 70 mm Hg
System
Specifications
Classification
Class II; IPXO, No AP/APG; Type BF
Applied part
Mattress
Power rating
~230 V AC/50 Hz
1A , 12 W
Fuse rating
250 V, T 1 A
Operating environment
Temperature
Use: 10° ~ 35° C
Storage: -15° ~ 50° C
Transport: -15° ~ 70° C
Humidity
Use: 20% ~ 80% without condensation
Storage: 10% ~ 90% without condensation
Safety standard
CE-marked
w w w.jar ven.se
B
C
E
D
A
The power switch is located on the right side of the pump. Press
“On/Off”
to start or stop the pump.
READY
Inflating the mattress takes approximately 15 minutes. When the lamp is lit, the
mattress is ready to use.
UNLOCK
The panel can be locked to prevent unauthorised people from changing the system
settings. To unlock the panel, press and hold the button for 5 seconds.
WEIGHT SETTING
Use the plus and minus arrows to select the weight setting that best suits the patient.
CARE MODE
This button provides maximum static pressure to facilitate nursing care procedures.
After 30 minutes the mattress will automatically return to the previous setting in use
before the care mode was selected.
ALARM
If the alarm sounds, press this button to mute and reset. The adjacent “ ” lamp will
light up if the air pressure in the mattress is too low. Check if necessary for any air leakage
(see page 12). The adjacent “ ” lamp will light up in the event of a power failure.
PUMP OVERVIEW – OC-15
15