Jamr Technology W02 User Manual Download Page 1

1

CONTACT INFORMATION

Wrist Type Blood Pressure Monitor is legal manufactured by:
Company name: Shenzhen Jamr Technology Co., Ltd
Address: A101-301,D101-201, Jamr Science & Technology Park, No. 2 Guiyuan Road,
Guixiang Community,Guanlan Street, Longhua District, Shenzhen 518100,PEOPLE'S
REPUBLIC OF CHINA
Website: www.jiemeirui.com
Telephone: 0086 755 85292057

Authorized European Representative:

Company Name: Shanghai International Holding Corp. GmbH (Europe)
Contact person: Mr JIN LIANG
Address:

Eiffestrasse 80, 20537 Hamburg, Germany.

Telephone:

+49-40-2513175 / 2513178

FAX:

+49-40-255726

E-mail:

[email protected]

Release date: 2021.03.18, Version:B1

Summary of Contents for W02

Page 1: ...ty Guanlan Street Longhua District Shenzhen 518100 PEOPLE S REPUBLIC OF CHINA Website www jiemeirui com Telephone 0086 755 85292057 Authorized European Representative Company Name Shanghai Internation...

Page 2: ...2 Wrist Type Blood Pressure Monitor Model Number W02 USER MANUAL...

Page 3: ...essure and its Measurement 5 3 Components of your blood pressure monitor 7 4 Using your Monitor for the First Time 8 5 Measurement Procedure 9 6 Care and Maintenance 16 7 Warranty 16 8 Certifications...

Page 4: ...is recommended that your physician review your procedure for using this device Blood pressure readings obtained by this device should be verified before prescribing or making adjustments to any medica...

Page 5: ...not suitable for use in the presence of flammable anesthetic mixtures with air or with oxygen or nitrous oxide Warning Not suitable for use in the presence of flammable agents Warning If having a pote...

Page 6: ...Note The cuff is treated as the applied part The user should contact the manufacturer for assistance if needed in setting up using or maintaining the device 2 Important Information on Blood Pressure a...

Page 7: ...ts of your measurements to independently alter the drug doses prescribed by your physician Further information If your values are mostly normal under resting conditions but exceptionally high under co...

Page 8: ...ood pressure 3 3 Features of Model W02 1 Talking function 2 Double users 2 x 120 sets memory 3 Date time display 4 Irregular heartbeat checking 5 Average value function 6 Low battery display 7 WHO fun...

Page 9: ...decide to delete the all record please keep the record in another way in case you need it some days later Take the battery out won t lead to a record missing 5 Measurement Procedure Note You should a...

Page 10: ...e facing you Apply the cuff so that the display is facing you b The distance between the cuff and the hand should be 1cm c Secure the cuff with the hook and loop adhesive so that it lies comfortably a...

Page 11: ...the device off If no button is pressed for 60seconds the device switches off automatically Recommended Use Methods 1 Recommendation that the PATIENT relax as much as possible and not talk during the...

Page 12: ...placement Replace both batteries with new ones Check the battery installation for proper placement of the battery polarities Er1 Sensor abnormal Check if the pump is working or not If it is working t...

Page 13: ...appear in this manual on the Blood Pressure Monitor W02 or on it s accessories Some of the symbols represent standards and compliances associated with the Blood Pressure Monitor W02 and its use Autho...

Page 14: ...the stored values With the unit off press the Memory button The display first shows A then shows an average of all measurements stored in the unit Please note Measurements for each user are averaged a...

Page 15: ...the monitor is not going to be used for a prolonged period the batteries should be removed Using rechargeable batteries You can also operate this instrument using rechargeable batteries Only use NiMH...

Page 16: ...y test Sensitive measuring devices must be checked for accuracy from time to time We recommend a periodical inspection of your unit by an authorized dealer every 1 years Please turn to local distribut...

Page 17: ...s 1 5 V Accessories 13 5 19 5cm Automatically power off 60 seconds Users Adult Software Version V1 1 1 Expected service life of the device and accessories 5 years Technical alterations reserved 10 EMC...

Page 18: ...rnal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emission CISPR 11 Class B The device is suitable for use in all est...

Page 19: ...EC 61000 4 5 1 kV line s to line s 2 kV line s to earth Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage v...

Page 20: ...ccording to the transmitter manufacturer and d is the recommended separation distance Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the...

Page 21: ...quency of transmitter m 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 7 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum outp...

Page 22: ...sing higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following e...

Page 23: ...23...