Introduction 

 

Xonic 100L Clamp-On Ultrasonic Flowmeter is fully digitalized, the state-
of-art  flowmeter  that  use  DSP  (digital  signal  processing)  technology  to 
measure  time  difference  of  ultrasonic  signals.  Please  read  this  manual 
carefully before installation and operation to ensure the best performance. 

 

 

 

Contents in this manual is subject to change by the manufacturer without 
any prior notice to the user.   

 

Safety Consideration 

Xonic 100L uses external power of AC110~220V . Follow all electrical and 
electronic safety rules to prevent any safety accident or damage of the flow 
computer during installation. Most applications are near to moisture, thus, 
be careful of any electric shock.   

 

Installation Steps   

Select best pipe installation point for transducer. 
1.  Input PIPE DATA. 
2.  Check the SENSOR DISTANCE displayed in flow computer. 
3.  Select the sensor INSTALLATION PLACE. 
4.  Mount the SENSOR horizontally to the pipe. 
5.  Install FLOW COMPUTER. 
6.  Connect the sensor and the flow computer via cable. 
7.  Supply POWER to the flow computer and turn the switch ON.   
8.  Input necessary DATA such as output and relay.   

 
Specifications 
 

Type 

Clamp-On Ultrasonic Flowmeter 

Principle 

Transit-Time 

    Pipe Size 

12 ~ 6000 mm 

Accuracy 

±1.0 % (single path), ±0.5 % (dual path) 

Flow Velocity 

±0.02 ~ 12.0 m/s 

Turn-Down Ratio 

500:1 

Repeatability 

0.25% 

Required Straight 
Run 

Upstream 10D, Downstream 5D 
(single path) 
Upstream 5D, Downstream 3D 
(dual path) 

Data Input 

Two 4~20mADC 

Data OUTPUT 

Two 4~20mADC for flow 
Relay for Total 
RS-232C / 485 Modbus 

Data Logger 

32 Mbytes (above 1.000,000 times) 

Display 

Graphic Color LCD 
(Flow, Total, Velocity, Delta T, 
Oscilloscope Shape) 

Temperature 

Flow Computer -20 ~ +75 

 

Transducer        -40 ~ +120 

 

Power 

110 ~ 220 VAC (12 ~ 24VDC option) 

Enclosure 

IP65 

Transducer 

IP68, submersible 

 

Summary of Contents for XONIC 100L

Page 1: ...Industrial Defense Instruments XONIC 100L ULTRASONIC TRANSIT TIME FLOWMETER FOR LIQUID FLOW User Manual 2021 www jain co kr Tel 82 2 856 4114 Fax 82 2 856 9503...

Page 2: ...ons Key Functions Understanding Display Best Transducer Installation Place Section 1 General SYSTEM LOCK SITE NAME APPLICATION FLANGE SET Section 2 CH SELECT Section 3 PIPE Data Section 4 LIQUID Type...

Page 3: ...Xonic 100L 2021 3 Delivery Items Flow Computer Transducer Sensor cable Mounting Track...

Page 4: ...Input PIPE DATA 2 Check the SENSOR DISTANCE displayed in flow computer 3 Select the sensor INSTALLATION PLACE 4 Mount the SENSOR horizontally to the pipe 5 Install FLOW COMPUTER 6 Connect the sensor...

Page 5: ...KEYS Keys Functions Press to enter the menu or back to the main display Press to enter the selected menu or save the input data Press to delete the text or number Use to input numbers Use to input dec...

Page 6: ...cted by user FLOW measured flow TOTAL total flow TOTAL UNIT total flow unit Best Transducer Installation Place Choose an installation location that has enough straight pipe runs 10D for upstream and 5...

Page 7: ...2 pairs of transducers DUAL CH 1 2 mode is used to measure two independent pipes separately and shows total flow by summing up CH 1 and CH 2 DUAL CH 1 2 mode is used to measure two independent pipes s...

Page 8: ...e data in other menus In case of single channel select CHANNEL 1 In case of dual channel user have to setup the data of each channel separately MAIN MENU 1 GENERAL 2 CH SELECT 3 PIPE 4 LIQUID 5 INSTAL...

Page 9: ...ALOG 7 LINING SONIC Vs 1 Metric mm 0 DIAG 8 LINING THICKNESS 2 US units inch PIPE 1 CARBON STEEL 2 DUCTILE_IRON 1 PIPEUNIT 3 CAST_IRON 2 PIPEMATERIAL 4 SUS 3 PIPESONIC Vs 5 ALUMINUM 4 PIPEDIAMETER 6 C...

Page 10: ...IT 3 TAR_EPOXY 2 PIPE MATERIAL 4 TEFLON 3 PIPE SONIC Vs 5 POLYETHYLENE 4 PIPE DIAMETER 6 ENAMEL 5 PIPE THICKNESS 7 GLASS 6 LINING MATERIAL 8 PLASTIC 7 LINING SONIC Vs 9 RUBBER 8 LINING THICKNESS 0 ASV...

Page 11: ...does not need to set flowmeter automatically displays viscosity of selected liquid Density User does not need to set flowmeter automatically displays density of selected liquid MAIN MENU 1 GENERAL 1 W...

Page 12: ...ting Type Normally CLAMP ON V is best for most applications V method is more accurate and easy to install V method means installation mode which mount two transducers onto one pipe side as per below M...

Page 13: ...orrosion inside please use Z MODE installation These pipes can sometimes make the ultrasonic signal very weak so flowmeter cannot work in V mode Also in the case that liquid is not clean for example w...

Page 14: ...area on the pipe using ruler Remember that the area must be large enough for Mounting Track installation with the sensor placement approximately in the center of the track Remove Pipe Coating and mak...

Page 15: ...c signal for flow measurement User simply select YES then Xonic 100 start AR mode automatic installation procedure User can see how AR mode find best signal Space INSTALL INFO Pipe CARBON STEEL OD 0 0...

Page 16: ...after clean pipe Signal Shape must be high in the middle area If signal shape is not high in the middle area please check pipe size pipe material etc Actual Zero This menu is useful when user can stop...

Page 17: ...ault value is 5 seconds This means display flow is average flow for 5 seconds MAIN MENU OPERATE 1 GENERAL 1 UPPER FLOW LIMIT 2 CH SELECT 2 LOWER FLOW LIMIT 3 PIPE 3 DEAD ZONE 4 LIQUID 4 FLOW AVERAGETI...

Page 18: ...LEAGC 9 DAMPING 0 FIX RISC FLOW AVERAGETIME sec 10 OPERATE 1 UPPER FLOW LIMIT 2 LOWER FLOW LIMIT 3 DEAD ZONE 4 FLOW AVERAGETIME 5 TOTAL FLOW SET 6 ALARM 7 CALIBRATION 8 ENABLEAGC 9 DAMPING 1 POSITIVES...

Page 19: ...and Kc is 0 99 then flow became 99 Enable Agc AGC is Automatic Gain Control function Enable is default 0 0000 0 000 0 0000 0 000 0 0000 0 000 0 0000 0 000 0 0000 0 000 0 0000 0 000 0 0000 0 000 1 MET...

Page 20: ...ENABLE OPERATE 1 UPPER FLOW LIMIT 2 LOWER FLOW LIMIT 3 DEAD ZONE DAMPING 4 FLOW AVERAGETIME 1 DISABLE 5 TOTAL FLOW SET 2 30 SEC 6 ALARM 3 1 MIN 7 CALIBRATION 4 5 MIN 8 ENABLEAGC 5 10 MIN 9 DAMPING 6...

Page 21: ...OLUTION FLOW VOLUMEUNIT 5 INSTALL 4 FLOW SCALE 1 CUBIC METER m 6 OPERATE 5 TOTAL VOLUMEUNIT 2 LITER L 7 FLOW 6 TOTAL RESOLUTION 3 GALLONS G 8 IN OUTPUT 7 TOTAL SCALE 4 CUBIC INCH CUI 9 DATALOG 8 BATCH...

Page 22: ...OTAL SCALE 8 BATCH TOTAL 1 NONE 9 TOTALIZER MODE 2 KILO FLOW FLOW SCALE 1 FLOW VOLUMEUNIT 2 FLOW TIMEUNIT 3 FLOW RESOLUTION 4 FLOW SCALE 1 CUBIC METER m 5 TOTAL VOLUMEUNIT 2 LITER l 6 TOTAL RESOLUTION...

Page 23: ...only totalizer negative flow If positive flow is 100 and negative flow is 10 then total is 10 1 FLOW VOLUMEUNIT 2 FLOW TIMEUNIT 3 FLOW RESOLUTION 4 FLOW SCALE 5 TOTAL VOLUMEUNIT 6 TOTAL RESOLUTION 7 T...

Page 24: ...ressure temperature user can use this function Just set MIN and MAX input SPAN the flowmeter sends the ANALOG INPUT data through RS 232C MAIN MENU 1 GENERAL 2 CH SELECT 3 PIPE ANALOG OUT 1 4 LIQUID IN...

Page 25: ...X INPUT SPAN is 10 Analog In 2 Same as ANALOG IN 1 1 SET ENABLE 1 ANALOG OUT 1 2 CALIBRATION_MIN 2 ANALOG OUT 2 3 CALIBRATION_MAX 3 RELAY OUT 1 4 MIN INPUT SPAN 4 RELAY OUT 2 5 MAX INPUT SPAN 5 ANALOG...

Page 26: ...a logger MAIN MENU 1 GENERAL 2 CH SELECT 3 PIPE 4 LIQUID 5 INSTALL 6 OPERATE DATALOG 7 FLOW 1 TIMESET 8 IN OUTPUT 2 RS 232C TIMESET 9 DATALOG 3 MEMORY YY MM DD HH MM SS 0 DIAG 4 WAVEDATA _ _ _ _ _ _ R...

Page 27: ...ressing ENT Format User can add and list the data here so the data will be download sequentially 1 BAUD RATE 2 DATA BITS 3 PARITY 4 STOPBITS 5 LINEFEED 1 1Bits 6 NETWORK ID 2 2Bits STOPBITS CONFIGURE...

Page 28: ...INFO FORMAT HFTA H Header S Site name N Channel No D Date C Time F Flow T Total I1 I2 AnalogIn1 2 U Unit A Alarm V Velocity RS 232C 1 CONFIGURE 2 HEADER 3 FORMAT 4 SEPERATOR 5 LOG INTERVAL 6 LOG TIMES...

Page 29: ...OG ENABLE RS 232C 1 CONFIGURE 2 HEADER 3 FORMAT 4 SEPERATOR 5 LOG INTERVAL 6 LOG TIMESYNC 7 LOG ENABLE 1 NORMAL 8 COMM MODE 2 Jain View 9 SELECT COMM 3 CDMA COMM 0 COMM INFO 4 MODBUS RTU COMM MODE RS...

Page 30: ...a logger 1 LOG OUTPUT 2 FORMAT 3 LOG INTERVAL 1 TIMESET MEMORY 4 LOG TIMESYNC 2 RS 232C 1 INTERNAL MEMORY 5 LOG ENABLE LOG OUTPUT 3 MEMORY 2 EXTERNAL MEMORY 6 PORT SELECT 1 LOG DATE 4 WAVEDATA 3 MEM L...

Page 31: ...PUT 2 FORMAT 3 LOG INTERVAL 4 LOG TIMESYNC 5 LOG ENABLE PORT SELECT 6 PORT SELECT 1 RS232 PORT 7 MEMORY CLEAR 2 FIRMWAREPORT MEMORY 1 LOG OUTPUT 2 FORMAT 3 LOG INTERVAL 4 LOG TIMESYNC 5 LOG ENABLE MEM...

Page 32: ...N LEVEL is amplitude level of signal The number is automatically set by flowmeter HOLD FLOW is functions to match with remote indicator When HOLD FLOW is 0 Xonic 100 send 4mA output When HOLD FLOW is...

Page 33: ...Xonic 100L 2021 33 DRAWINGS...

Page 34: ...onic 100L 2021 34 208 250 260 282 95 MOUNTING HOLE PATTERN 6 HOLE 150 260 Ultrasonic Flowmeter Xonic 100 Flow computer Outline 27 208 96 3 TRANSDUCER CABLES POWER IN FLOW COMPUTER MODEL ID LABEL LATCH...

Page 35: ...Xonic 100L 2021 35...

Page 36: ...Xonic 100L 2021 36...

Page 37: ...Xonic 100L 2021 37 A A B C D 23 42 37 63 15 80 35 65 40 72 50 250 35 93 50 86 200 500 50 145 76 111 500 6000 LTB LTC LTD LTE mm Dimension B D c P N...

Page 38: ...onic 100L 2021 38 REFLECT MODE INSTALLATION 102 TP TO FLOW COMPUTER UP STRAP GUIDE TRANSDUCER MOUNTING TRACK SERIES CUSTOMER S PIPE CABLE FLOW TRANSDUCERS INSTALLATION AND MOUNTING TRACKS REFLECT MODE...

Page 39: ...S INSTALLATION AND MOUNTING TRACKS DIRECT MODE DIRECT MODE INSTALLATION TO FLOW COMPUTER UP TRANSDUCER CLAMP STRAP GUIDE TRANSDUCER SET PAIR MOUNTING TRACK SERIES CUSTOMER S PIPE CABLE FLOW CABLE TO F...

Page 40: ...Xonic 100L 2021 40 INSTALLATION OUTLINE 102 S SF ADJUSTABLE MOUNTING STRAP...

Page 41: ...Xonic 100L 2021 41 USE OF COUPLANT COMPOUND TRANSDUCERS INSTALLATION 102 C COUPANT APPLY A THIN BAND OF COUPLANT TP BASE OF EACH TRANSDUCER BEFORE INSTALLING ON PIPE...

Page 42: ...Xonic 100L 2021 42 710 55 Digital ro 33 gil Guro gu Seoul Korea TEL 82 2 856 4114 FAX 82 2 856 9503 www jain co kr...

Reviews: