SP-5000M-PMCL / SP-5000C-PMCL
- 79 -
2.4.3 Device Control
Name
Interface
Acce
ss
Short
ASCII
Values
MIN
MAX
DEFA
ULT
Description
DeviceFirmwar
e
Version
I String
R/O
VN
Firm Ver. No.
-
-
-
VN?<CR><LF>
DeviceReset
I
Comman
d
W/O CRS00 1
-
-
-
CRS00=1<CR><LF>
2.4.4 Image Format Control
Name
Interface
Acce
ss
Short
ASCII
Values
MIN
MAX
DEFA
ULT
Description
Height
I Integer
R/W HTL
Min
~
(Max - OffsetY)
1(Mon
o)
2(Bay
er)
2048
2048
HTL=[Param.]<CR><LF
>
HTL?<CR><LF>
Width
I Integer
R/W WTC
Min
~
(Max - OffsetX)
16
2560
2560
WTC=[Param.]<CR><LF
>
WTC?<CR><LF>
Offset Y
I Integer
R/W OFL
Min
~
(Max – Height)
0
2047
(Mono
)
2046
(Baye
r)
0
OFL=[Param.]<CR><LF
>
OFL?<CR><LF>
Offset X
I Integer
R/W OFC
Min
~
(Max – Width)
0
2544
0
OFC=[Param.]<CR><LF
>
OFC?<CR><LF>
BinningHorizont
al
I Integer
R/W HB
1: Normal / 2: Binning
mode
1
2
1
HB=[Param.]<CR><LF>
HB?<CR><LF>
only Mono
BinningVertical
I Integer
R/W VB
1: Normal / 2: Binning
mode
1
2
1
VB=[Param.]<CR><LF>
VB?<CR><LF>
only Mono
PixelFormat
I
Enumera
tion
R/(
W)
BA
Mono model:
0: Mono8
1: Mono10
2: Mono12
Bayer model:
0: BayerRG8
1: BayerRG10
2: BayerRG12
0
2
0
BA=[Param.]<CR><LF>
BA?<CR><LF>
TestImageSelec
tor
I
Enumera
tion
R/W TPN
0: Off
1:
GreyHorizontalRamp
2: GreyVerticalRamp
3:GreyHorizontal
RampMoving
4: Horizontal Colorbar*
5: Vertical Colorbar*
6: Moving Colorbar*
(* Bayer model
only)
0
6
0
TPN=[Param.]<CR><LF
>
TPN?<CR><LF>