3
Err-03
When calibration, the placed weight is too small and it refuses to calibrate.
(Weight should more than 10% max. capacity)
- - - -
Overload, show this as the weight over 100%FS+9e, with the DING sound.
Then you should reduce the weight.
4. Operations
4-1
Power on and Power off:
①
Power on:
Press <ON/ZERO> key, It displays software version after “DING” sound, and then dc-x.xx (voltage of battery
is 3~4.2V), self-inspection, 9,8,7···1,0, and the “zero” lamp lighting. Now it’s in working state.
②
Power off:
Press <OFF> key to power off.
③
Auto power off:
It will be auto power off after 10 minutes when scale is in at ZERO and with no operation.
④
Low voltage off:
It will be auto power off when the voltage is too low.
4-2 Calibration
1. Switch on the adjust key
(
the adjust key is in the indicator
)
.
Press <ON/ZERO> to power on .
2. Press <SET/0~9> and <ON/ZERO> 3 seconds at the same time.
It shows the last capacity. For example: c-60 means 60kg.
Press <SET/0~9> to choose c-30, c-60,
c-100, c-150,c-200, c-300,c-600,
c-1500, c-3000, c-6000
(
c is capacity and unit is kg
)
3. Press <TARE> to do tare action.
4. Now it displays internal number 0. If it is not 0, press <ON/ZERO> to zero.
5. Place more than 10% max. capacity standard weights on plate.
Press <TARE> to input the value of weights after internal number is stable.
Press <SET/0~9> to add 1. Press <TARE> to input next value till to finish.
And then it shows CAEnd for 1 second and return to weighing.
(Please attention to the place of decimal point when inputting values)
6. If display Err03, it means calibration fault.