    JWI-501 

 

 

3-9-5

   

Zero

: set auto zero.   

Default = 1e

 

When it displays 

ZEro

,  press <

TARE

>to enter. It displays 

ZEro  x.x

(x.x is auto zero 

range

0.5d

1d

,

  1.5d

,

  2d

,

  2.5d

,

  3d

,

  3.5d

,

  4d

,

  4.5d

  and 

5d

d

  is division value when 

n=3000). Press <

SET/0~9

> to choose. Press <

TARE

> to confirm and return weighing. 

 

3-9-6 

 

bUZZEr

: set buzzle.  

Default = off

 

When it displays 

bUZZEr

, press <

TARE

> to enter. Press <

SET/0~9

> to choose 

on

 or 

off

on

 is using sound in upper and lower limit situation; 

off

 is without sound. 

 
3-9-7

   

d

: choose division.  

Default = 3000 division value

 

  When it displays 

d

press  <

TARE

> to enter. Press  <

SET/0~9

> to choose division. 

Press <

TARE

> to confirm, and there have three divisions of each capacity. When you 

choose the little one, the speed of display will be slower. 

 
3-9-8  

on-rA

: powering on range.   

Default=100

 

When it displays 

on-rA

, press <

TARE

> to enter. It will show 

100 

which

 

indicates 

that no limit is imposed on the powering on range. In case the zero point upon 

powering on is greater than 20% of the maximal weighing capacity, Err-0  and 

Err-04 will be displayed upon powering on, and the tare should be reduced to less 

than 20% of the maximal weighing capacity.   

 
3-10

 

Brightness

 

Press and hold <

ON/ZERO

>. it displays 

lc-

. Press <

SET/0~9

> to choose 

lu-1

lu-2

.   

lu-1 

is save power mode (Recommend). Press <

TARE

> to confirm. 

Default = lu-1 

 

4 Calibration 

1

  Press <

ON/ZERO

> to turn on. After self-inspection, it comes into normal weighing 

mode (no wind and quaking) and under zero;   

2

  Press <

SET/0~9

>  together with <

ON/ZERO

>  for  3 seconds

it shows the last 

capacity. Press <

SET/0~9

> to choose 

c-30

c-60

c-150

c-300

c-600

c-1500

c-3000

c-6000

c

 is capacity

unit is kg

3

  Press <

TARE

> to confirm; 

4

  Now it will show the offset value “

0

”. If it is not 0, press <

ON/ZERO

> to zero; 

5.  Place some standard weights on plate. After offset value is stable, press <

TARE

> to   

Summary of Contents for JWI-501 Series

Page 1: ... 1 2 3 Operation guide 3 1 Install and adjust 3 2 Turn on and turn off 3 3 Zero 3 4 Weighing 3 5 Tare weighing 3 6 Save power function 3 7 Battery indicate 3 8 Charge 3 9 Set and use parameter 3 10 Brightness 2 2 2 2 2 3 3 3 3 3 5 4 Calibration 5 5 Notes 5 1 Normal note 5 2 For sealed lead acid battery 5 5 6 ...

Page 2: ...several national patents for the scale The waterproof and dustproof grade is up to IP68 Please read the instruction manual before usage 1 Main technical index Max Capacity 30kg 60kg 150kg 300kg 600kg Min Capacity 20e Test division e 10g 20g 50g 100g 200g Display division d1 5g 10g 20g 50g 100g Display division d2 2g 5g 10g 20g 50g Max tare 100 max capacity Accuracy III Working temperature 5 35 Sto...

Page 3: ...ews Adjust feet to let the platform in level 3 2 Turn on and turn off Turn on Press ON ZERO key It displays software version After that it will show dc x xx voltage of battery Self inspection starts After countdown the zero lamp will light Now the scale is in a working state Turn off Press OFF key Auto turn off When scale is in auto off state It will be auto off after 10 minutes of no operation se...

Page 4: ...lower than 6v power will exhaust 3 8 Charge Power supplied by an internal rechargeable lead acid battery 6v 4 5ah When it shows batlo please charge immediately The time of full charging is 12 hours While charging no need to turn the scale on 3 9 Set and use parameter Press SET 0 9 for 3 seconds it will enter setting mode and show the main menu Press SET 0 9 again it displays rAngE set limit UnItS ...

Page 5: ...9 2 2 Counting When it displays pcs press TARE Now pcs lamp lighting Press and hold ON ZERO it displays count and return to 0 Then put the sample as much as possible but don t more than max capacity After stable press TARE Press SET 0 9 to add 1 and press TARE to choose After finishing it displays pcs and pcs lamp lighting exit counting choose a weighing unit Note Quantities of max less 30000 3 9 ...

Page 6: ...s on rA press TARE to enter It will show 100 which indicates that no limit is imposed on the powering on range In case the zero point upon powering on is greater than 20 of the maximal weighing capacity Err 0 and Err 04 will be displayed upon powering on and the tare should be reduced to less than 20 of the maximal weighing capacity 3 10 Brightness Press and hold ON ZERO it displays lc Press SET 0...

Page 7: ...mage due to overload is not within the range of guaranteed maintenance Don t touch chemicals such like solvent etc Keep the scale clean in order to weighing well There is a lead seal on the bottom of the scale Users are not allowed to remove the seal If the seal is damaged the scale is not guaranteed for maintenance 5 2 For sealed lead acid battery The working temperature of battery is 10 40 It ca...

Page 8: ...JWI 501 7 ...

Reviews: