JWI-3000 10
stable
:
Stable print
(
weight is equal or more than 20d
)
key
:
Manual print by pressing key
14.
Print format
See the appendix (more than 100 formats. The appendix just shows two formats.)
15.
Filtering setting
Set the filtering level in which the stable indication turns on. The higher the setting, the slower
stabilization time
Options:
1
,
2
,
3
,
4
16.
Tare/Zero condition
stable
:
Only after the stable indication appears, Tare/Zero function acts after pressing
down key
TARE
or
ZERO
always
:
Tare/Zero function acts by pressing down key
TARE
or
ZERO
even if it is not
stable
auto:
Press down key
TARE
or
ZERO
even if it is not stable, but Tare/Zero function acts
after stable
17.
weight memory parameters.
ON
display the previous weight when powering on again
OFF
not display the previous weight when powering on again
18.
parameter initialization
Press
UNIT/SET
twice to begin initialization until the window displays
“
OK”
6 Calibration
Note:
Before calibration, please set the capacity first. The unit used in calibration is the one that
has been set before. During the calibration procedure, LONG press
ZERO/ESC
to return to
normal weighing mode without saving.
Here we take 3kg/10g as an example
1. Press and hold
TARE/
while powering on. Do not release it till the window displays
“
CAL”
2. With no load on the weighing pan, press
TARE/
to start zero point calibration.
3. Wait till the window displays the first calibration value. (the window displays
1.000kg
)
Note:
The first calibration value is default. If the capacity has been changed, the default
value is 1/3 of full load.