C I N E B R A T E User Manual
21
Data Code
Description
114
status, Software Version
Output Format
: qqqqq,vvvvv CRLF
where
: vvvvv = Software version
Output Example:
00000,3.09
115
status, Battery Status
Output Format
: qqqqq,v CRLF
where
: v = 0, Battery OK
v =1, Battery LOW
Output Example:
00000,0
116
status, Accessory List
Output Format
: qqqqq,
nn,ss,tt,pp,rr
CRLF
where
: nn = ID number of Accessory
ss = Accessory Name
tt = Accessory Type
pp = Photometry Mode - Possibilities are: Luminance, Illuminance,
Luminous Intensity, or Luminous Flux
rr = Radiometry Mode - Possibilities are: Radiance Irradiance Radiant
Intensity or Radiant Flux
Output Example:
00000,FF75,Primary,Luminance,Radiance
TABLE 4 - REMOTE CONTROL DATA CODE SUMMARY
R
EMOTE
C
ONTROL
E
RROR
C
ODES
R
EMOTE
C
ONTROL
M
EASUREMENT
E
RRORS
Error
Meaning
-1
Light source not constant.
-2
Light overload
–
signal too intense.
-3
Cannot Sync to light source. Light source frequency below 20Hz, above 400
Hz or signal too low to Sync.
-4
Adaptive mode error.
-8
Weak light
–
insufficient signal.
-9
Sync Error.
-10
Cannot Auto Sync to light source.
-12
Adaptive mode time out. Light source not constant.
TABLE 5 - REMOTE CONTROL MEASUREMENT ERRORS