104
©2023 J.A. Woollam Co.
M-2000® Hardware Manual
“Use Separate Angle Offsets” may also be enabled (if applicable) – see AOI
Correction section of this manual. Check ‘Show Advanced Config. Options’ to see
the window effects and delta offsets displayed. DO NOT edit these numbers
manually. These are calculated after the alignment is performed. The
configuration parameters should not change and do not need to be checked
each time the focusing probes are added.
Figure 6-2. Hardware Configuration window.
Caution: The focusing lenses and their coatings can easily be damaged. Be
careful when handling the lens mounts so that nothing touches the
lenses.
6.2
Installation
Focusing optics should not be installed during the initial Alignment and
Calibration Procedures described earlier in this manual. If the system is ready for
general use, the focusing optics can be installed using the following procedure.
Caution: Remove any sample positioning pins from the X axis of the sample
stage prior to attaching the focusing probes. These pins may
prevent proper alignment of the focusing optics or cause damage to
the focusing tilt stages due to improper strain.
1.
Select
Hardware
>
Controls
>
System
>‘Install Focus Probes’.