– Fill the pressure chamber with approx. 0.5 l water and
dissolve the amount of deliming agent recommended
in the corresponding instructions for use.
– Close and bolt the pressure chamber lid.
– Set the thermostat to 80 °C and the timer to
5 minutes.
– Once the program is completed, open the lid and
check the cleaning results.
– Remove any large, loose particles from the chamber.
– If necessary, repeat the procedure until the heating
element and pressure chamber are clean.
– After decalcifying, repeat the procedure with fresh
water to rinse the pressure chamber and the outflow
system. (Once the operating pressure is reached, the
program can be interrupted by setting the timer to 0)
– Empty the return water canister and rinse with fresh
water.
Deliming agents are caustic and toxic.
Avoid contact with skin and eyes. In case
of accidental contact, rinse with copious
amounts of fresh water.
Condensation accumulated in the sight glass (1a) can be
drained off by pushing the release button (1b).
At three-months intervals, unscrew the canister cap (19)
and clean (clear) the small air hole (25) of the canister
hose (21). Access to the hole (25) is gained by loosening
the screw (26) and removing the collar (24).
8.1 Power failure or blown fuse
– In the event of a power failure, the automatic program
is interrupted. Once the power supply returns, the
program will automatically continue.
– To take the objects out of the pressure chamber
during prolonged power failures, proceed as described
under 3a.
– If the fuse (15) of the apparatus is defective, open the
fuse cap and replace the fuse.
If problems arise, which are not mentioned in the
table below, please contact our Customer Service
Department.
8.2 Repair
Repairs may only be carried out by a certified Ivoclar
Vivadent Service Center. Please contact our Customer
Service Department.
12
8. What If …
IvomatIP3-5sp-sb.qxd 01.07.2005 16:12 Uhr Seite 12