7. Procedure in case of repair
8. Product specifications
The warranty period for Bluephase Style M8 is 2 years
from the date of purchase (battery 1 year). Malfunctions
resulting from faulty material or manufacturing errors
are repaired free of charge during the warranty period.
The warranty does not provide the right to recover any
material or non-material damage other than the ones
mentioned. The apparatus must only be used for the
intended purposes. Any other uses are contraindicated.
The manufacturer does not accept any liability resulting
from misuse and warranty claims cannot be accepted in
such cases. This is particularly valid for:
• damage resulting from improper handling, especially
incorrectly stored batteries (see chapter Product speci-
fications: Transportation and storage conditions);
• damage to components resulting from wear under
standard operating conditions (e.g. battery);
• damage resulting from external influences, e.g. blows,
drop to the floor;
• damage resulting from incorrect set-up or installation;
• damage resulting from connecting the unit to a power
supply whose voltage and frequency do not comply
with the ones stated on the rating plate;
• damage resulting from improper repairs or modifica-
tions that have not been carried out by certified
centers.
In case of a claim under warranty, the complete appara-
tus (handpiece, charging base, power cord and power
pack) must be returned, carriage paid, to the dealer or
directly to Ivoclar Vivadent, together with the purchase
document. Use the original packaging with the corre-
sponding cardboard inserts for transportation. Repair
work may only be carried out by a certified Ivoclar
Vivadent Service Center. In case of a defect that cannot
be rectified, please contact your dealer or your local
Service Center (see addresses on the reverse side). A
clear description of the defect or the conditions under
which the defect occurred will facilitate locating the
problem. Please enclose this description when returning
the apparatus.
Light source
LED
Wavelength range
430–490 nm
Light intensity
800 mW/cm
2
± 10%
Operation
3 min. on / 7 min. off (intermittent)
Light probe
10 mm, black, disinfectable and
autoclavable
Anti-glare cones
autoclavable
Signal transmitter
acoustic after 10 seconds and every time
the start button / time selection button is
pressed
Dimensions of the
handpiece
(without light probe)
L = 180 mm W = 30 mm H = 30 mm
Weight of the handpiece
120 g (including battery and light probe)
Operating voltage
handpiece
3.7 VDC with battery
Operating voltage
Charging base
5 VDC
Power pack
Input: 100–240 VAC, 50-60 Hz 310 mA
Output: 5 VDC / 2 A
Manufacturer Friwo
Type: FW7401M/05
Weight of the power pack 165 g
Operating conditions
Tempe10 °C to +35 °C
(+50 °F to +95 °F)
Relative humidity 30% to 75%
Ambient pressure 700 hPa to 1060 hPa
Dimensions of the
charging base
D = 125 mm, H = 70 mm
Weight of the charging
base
195 g
charging time
approximately 2 hours (with the battery
empty)
Power supply of the
handpiece
Li-Po battery (approx. 20 min. with a new,
fully charged battery)
Transportation and stor-
age conditions
Temperature –20 °C to +60 °C
(–4 °F to +140 °F)
Relative humidity 10% to 75%
Ambient pressure 500 hPa to 1060 hPa
The curing light has to be stored in a
closed, roofed room and must not be
exposed to hard shocks.
Battery:
– Do not store at temperatures above
40 °C / 104 °F (or 60 °C / 140 °F for a
short period). Recommended storage
temperature 15–30 °C (59–86 °F).
– Always store the battery charged and
not for longer than 6 months.
Delivery form
1 Handpiece
1 Battery
1 Charging base with power cord and
power pack
1 Light probe 10 mm, black
3 Anti-glare cones
1 Set of Instructions for Use