6. What if …?
Indicator
Causes
Error rectification
Red “x”
lights up
The device is overheated.
Allow the device to cool down and try again after a
certain time. If the error persists, please contact your
dealer or your local Service Centre.
Electronic component of the
handpiece is defective.
Remove and reinsert the battery. If the error persits,
please contact your dealer or your local Service
Centre.
Red “x” and
battery symbol
light up
Battery empty
Place the device in the charging base and charge it.
Battery contacts dirty
Remove battery and clean the battery contacts.
3s flashes
In the 3sCure program, the curing time
is set to 3 seconds and cannot be
changed. The 3sCure program can be
used no more than twice in direct
succession.
After two consecutive curing cycles, an extended
waiting time of 30 seconds is necessary for safety
reasons before the next curing cycle can be started.
If for compelling reasons an additional curing cycle is
required before the prescribed interval of 30 seconds
has elapsed, the 3sCure program can be activated
again by selecting it anew with the program selection
button.
The charging base
is not illuminated
during charging
– Power pack not connected or
defective
– Battery fully charged
Check if the power pack is correctly positioned in
the charging base or if the power pack is connected
to the power supply by means of the power cord.