6.3.2 Control mode
The CM Tiny IQ has several options to control the output voltage and current. This allows the
generator to be set to VC mode (Voltage Control), CC mode (Current Control) or CLFB-mode
(Closed Loop FeedBack). By default, the CM Tiny IQ is set to VC mode (Figure 8).
Operation VC (Voltage Control) mode:
Figure 12, operation VC mode
If the CM Tiny IQ is set to VC mode, the desired output voltage must be set with the parameter
"Setpoint voltage." The generator will now consistently maintain the set output voltage and
adjust it depending on the load with a larger or smaller current. The charging current depends
on the load of the generator and the possible fouling of the charging bar/electrode. The
maximum output current can be set using the parameter "Output current limit". Only if the output
current reaches the value of the "Output current Limit" parameter, will the output voltage drop.
Operation CC (Current Control) mode:
Figure 13, operation CC mode
If the CM Tiny IQ is set to CC mode, the desired output current must be set with the parameter
"Setpoint current." The generator will now consistently maintain the set output current and
adjust it depending on the load with a larger or smaller voltage. The maximum output voltage
can be set using the parameter "Output voltage limit". Only if the output voltage reaches the
value of the "Output voltage limit" parameter, will the output current drop.
Summary of Contents for Simco-ION Charge Master Tiny
Page 22: ...CMTinyCMTinyIQ_UM_9752097710_NL_D_GB_F_V1_1 21 NL Reserveonderdelen ...
Page 45: ...CMTinyCMTinyIQ_UM_9752097710_NL_D_GB_F_V1_1 44 Ersatzteile ...
Page 67: ...CMTinyCMTinyIQ_UM_9752097710_NL_D_GB_F_V1_1 66 Spare parts ...
Page 89: ...CMTinyCMTinyIQ_UM_9752097710_NL_D_GB_F_V1_1 88 Pièces de rechange ...