Aerostat XC Heater 5100550-M Rev 7
19
4.7 Calibration
The Aerostat XC's ion output is inherently balanced by design, so
there are no calibration adjustments. If after checking the ion
balance as outlined above, an unbalance or offset voltage exists in
excess of 0±5 volts, contact Simco-Ion Technical Support at 510-
217-0470 or email [email protected].
Never open the case of the unit. This exposes hazardous voltages.
If repairs are necessary, call Simco-Ion for a Return Authorization
Number and return the unit to the factory.