background image

Operating manual VK 30.03 

 

Parameterization pyrometer PA 

The current output 2 of the pyrometer is deactivated by default. In order 
to  use  the  VK  30.03,  you  must  parameterize  the  analogue  output  2  as 
voltage input and assign it to the emissivity or emissivity ratio.  
In  the  following  configuration  example,  the  voltage  input  is  used  for  the 
external adjustment of the emissivity ratio.  
 

5.1 

Configuration I/O (configuration layer: 

C010

Parameter  Function 

Explanation 

AO2. 

Analogue output 2 

  OFF

 off 

AI.FN 

Analogue input 
function 

EPS.Q

 emissivity ratio

 

AI.U1 

Ain Scaling 

Define lower limit of voltage for input voltage 
 

(2 V)

 

AI.U2 

Ain Scaling 

Define upper limit of voltage for input voltage 
 

(10 V)

 

AI.V1 

Ain Scaling 

Input of lower voltage value  
(example 

100% 

emissivity ratio) 

AI.V2 

Ain Scaling 

Input of upper voltage value  
(example 

110% 

emissivity ratio) 

SAVE 

Save 

Save changes / exit menu 

 

5.2 

Display of internal measured values

 

According to standard the current temperature is shown in the display of 
the  pyrometer.  Should  the  current  value  at  the  voltage  input  is  to  be 
shown  in  the  display,  change  to  code  page  C020  and  select parameter 
AIN. 
 
Configuration layer: 

C020

 

Parameter  Function 

Explanation 

AIN 

Initial value at ana-
logue input 

Current value of analogue input 

 

Summary of Contents for VK 30.03

Page 1: ...Operating manual VK 30 03 Material selections switch VK 30 03 Ident No 109 8756 12 2018...

Page 2: ...suppression EMV 3 2 5 Quality Management Certification 3 2 6 Environmental Management 3 3 General 4 4 Connection 4 4 1 Connection plan 4 4 2 Current output 4 20 mA 4 4 3 Parameterization VK 30 03 6 5...

Page 3: ...and or material damage CAUTION This symbol indicates important information which if neglected might result in material selection switch damage malfunction or breakdown PLEASE NOTE This symbol points o...

Page 4: ...pproval of the Manufacturer are expressly prohibited by law Violations shall be subject to compensation claims by the Manufacturer The right to claim additional indemnities remains reserved 2 Safety T...

Page 5: ...ic compatibility European certification EN 61000 6 4 EN 61000 6 2 EN 55011 When connecting a power supply unit make sure that is also conforms to these standards Radio interference may arise if the py...

Page 6: ...tage required for operation must be supplied by a separate power supply This power supply unit must conform to directive DIN IEC 61010 3 GND current output 4 0 4 20 mA 12 Power supply 24 DC 13 Power s...

Page 7: ...Operating manual VK 30 03 5 Pyrometer PA PIN 3 GN green Analogue output 2 analogue input PIN 8 RD red L GND VK 30 03 Clamp 4 Current output Clamp 3 L GND...

Page 8: ...ues use the arrow keys and Configuration of the analogue output Use the keys and to select 4 20 mA Set point for power on rEStor set Point value remains valid in case of power failure Selection with t...

Page 9: ...Q emissivity ratio AI U1 Ain Scaling Define lower limit of voltage for input voltage 2 V AI U2 Ain Scaling Define upper limit of voltage for input voltage 10 V AI V1 Ain Scaling Input of lower voltag...

Page 10: ...g claims has expired you will no longer be able to make any claims for compen sation of damage or loss 6 2 Packaging The packages used by KELLER HCW are made of carefully selected environmentally comp...

Page 11: ...Operating manual VK 30 03 9...

Page 12: ......

Reviews: