Operating manual VK 30.03
7
5
Parameterization pyrometer PA
The current output 2 of the pyrometer is deactivated by default. In order
to use the VK 30.03, you must parameterize the analogue output 2 as
voltage input and assign it to the emissivity or emissivity ratio.
In the following configuration example, the voltage input is used for the
external adjustment of the emissivity ratio.
5.1
Configuration I/O (configuration layer:
C010
)
Parameter Function
Explanation
AO2.
Analogue output 2
OFF
off
AI.FN
Analogue input
function
EPS.Q
emissivity ratio
AI.U1
Ain Scaling
Define lower limit of voltage for input voltage
(2 V)
AI.U2
Ain Scaling
Define upper limit of voltage for input voltage
(10 V)
AI.V1
Ain Scaling
Input of lower voltage value
(example
100%
emissivity ratio)
AI.V2
Ain Scaling
Input of upper voltage value
(example
110%
emissivity ratio)
SAVE
Save
Save changes / exit menu
5.2
Display of internal measured values
According to standard the current temperature is shown in the display of
the pyrometer. Should the current value at the voltage input is to be
shown in the display, change to code page C020 and select parameter
AIN.
Configuration layer:
C020
Parameter Function
Explanation
AIN
Initial value at ana-
logue input
Current value of analogue input