Operating Steps: 

Must wear rubber insulation gloves, seat under the rubber insulation pad!Only when the test lamp is off 

and there is no high voltage output can the test object be connected or disassembled! 

 

1. Connect the object under test and make sure that the voltmeter indicates "0" and the test lamp is off.If 

it is AC test, select the AC high-voltage output hole connection test line; if it is DC test, select the DC 

high-voltage output hole connection test line and connect the ground line;2. Set the value required for 

leakage current test; 

1) press the preset/test switch; 

2) select the current range, 0 ~ 2mA range or 2mA ~ 10mA range; 

3) adjust the required leakage current alarm value; 

4) pop the preset/test switch.3. Manual test: 

1) set the timing switch to the off state. If it is AC test, the AC/DC switch will pop up. If it is DC test, press 

the AC/DC switch.Press the start button, the test light will be on, and turn the voltage adjustment button 

to the required indicating value; 

2) if the leakage current index of the measured object exceeds the specified leakage current value, the 

instrument will automatically cut off the output voltage, and the buzzer will give an alarm, and the 

over-leakage indicator light will be on. At this time, the leakage current index of the measured object is 

unqualified. Press the reset key to eliminate the alarm sound.On the contrary, the leakage current index 

of the measured object is qualified. 

 

  Timing test: 

1) press the time switch to adjust the value of the time timer and set the required test time value; 

2) set the alarm value of leakage current and press the start button to adjust the voltage to the required 

test value; 

3) if the timing time is up, the test voltage is cut off and the test lamp goes out, then the leakage current 

index of the tested object is qualified;If the current is too large, the over-leakage light will be on in the 

timing process and the buzzer will give an alarm. The leakage current index of the measured object is 

unqualified. Press reset 

Key to eliminate the alarm sound. 

 

Caution: 

1. The operator must wear rubber insulating gloves and put rubber insulating pads under the seat to 

prevent high-voltage electric shock; 

2. The instrument must be reliably grounded; 

Summary of Contents for LK2674A

Page 1: ......

Page 2: ...anual carefully before you use it We provide one year quality assurance and the consummation after sale service If during the using process you discover any question please promptly contact us Thanks...

Page 3: ...electrical appliances power switches wires and cables transformers wiring terminals high voltage bakelite electrical appliances motors medical treatment chemical industry instruments and meters as wel...

Page 4: ...larm value AC DC 0 3ma 2mA 2mA 10mA continuous setting 4 Time test range 1s 99s continuous setting and manual operation 5 Capacity 200VA 6 Output waveform sine wave AC DC 7 Power supply 198V 242V AC 4...

Page 5: ...indicating the breakdown of the measured object and over leakage is unqualified 7 High voltage output terminal DC high voltage output terminal 8 High voltage output terminal AC high voltage output ter...

Page 6: ...est switch to continuously set the alarm value of leakage current 0 3ma 2mA or 2mA 10mA 15 timing switch press the switch adjust the time timer can set the test time pop up the switch the timer does n...

Page 7: ...ing value 2 if the leakage current index of the measured object exceeds the specified leakage current value the instrument will automatically cut off the output voltage and the buzzer will give an ala...

Page 8: ...do not use or store the instrument in a high temperature damp and dusty environment 11 After the instrument has been used for one year it must be sent to the metrology department or back to the facto...

Page 9: ...www itokin2000 com Tel 02 9744354 6 Hotline 088 7777277 Email salesitokin gmail com 9...

Reviews: