Panel Instructions: 

1. Power switch; 

2. Start button: when pressed, the test lamp will be on, and the instrument will output high voltage; 

3. Reset button: when pressed, the test lamp will be off, and there will be no high-voltage output; 

4. Voltage regulation button: adjust the output voltage, which is small in inverse time and large in 

reverse; 

5. Test light: the light is on, indicating that the high voltage has been started, and the high voltage is 

disconnected when the light is off; 

6. Over-leakage light: the light is on, indicating the breakdown of the measured object, and over-leakage 

is unqualified; 

7. High voltage output terminal: -DC high voltage output terminal; 

8. High voltage output terminal: AC high voltage output terminal; 

9. Voltmeter: output voltage indicator; 

10, time timer: 1s~99s timing adjustment, you can set the required test time value; 

11. Selection switch of the range of leakage current: switch the range of leakage current indicating 

ammeter. The switch pops out as the range of 0-2ma, and the switch presses down as the range of 

2-10ma. 

12. Preset/test switch of leakage current: press the switch to set the alarm value of leakage current, and 

the switch will pop out. When normal, it will be the test state, and the value of leakage current can be 

detected in real time through "leakage current indicating ammeter". 

Summary of Contents for LK2674A

Page 1: ......

Page 2: ...anual carefully before you use it We provide one year quality assurance and the consummation after sale service If during the using process you discover any question please promptly contact us Thanks...

Page 3: ...electrical appliances power switches wires and cables transformers wiring terminals high voltage bakelite electrical appliances motors medical treatment chemical industry instruments and meters as wel...

Page 4: ...larm value AC DC 0 3ma 2mA 2mA 10mA continuous setting 4 Time test range 1s 99s continuous setting and manual operation 5 Capacity 200VA 6 Output waveform sine wave AC DC 7 Power supply 198V 242V AC 4...

Page 5: ...indicating the breakdown of the measured object and over leakage is unqualified 7 High voltage output terminal DC high voltage output terminal 8 High voltage output terminal AC high voltage output ter...

Page 6: ...est switch to continuously set the alarm value of leakage current 0 3ma 2mA or 2mA 10mA 15 timing switch press the switch adjust the time timer can set the test time pop up the switch the timer does n...

Page 7: ...ing value 2 if the leakage current index of the measured object exceeds the specified leakage current value the instrument will automatically cut off the output voltage and the buzzer will give an ala...

Page 8: ...do not use or store the instrument in a high temperature damp and dusty environment 11 After the instrument has been used for one year it must be sent to the metrology department or back to the facto...

Page 9: ...www itokin2000 com Tel 02 9744354 6 Hotline 088 7777277 Email salesitokin gmail com 9...

Reviews: