ITech IT6800A Series Programming Manual Download Page 51

 

 

Error Information 

 

Copyright © Itech Electronic Co., Ltd.

 

42

 

Chapter13  Error  Information 

Error Code List 

Error 
code 

Error information   

Example 

110 

"No input command" 

Send empty command. 

120 

"Parameter 
overflowed" 

The setting value exceeds the range. For 

example, CURRent 100.0 

130 

"Wrong 

units 

for 

parameter" 

For  example,  CURRent  5.0V,  the  unit  is 
wrong. 

140 

"Wrong 

type 

of 

parameter" 

For example, OUTPut:TIMer 100001.0. 

150 

"Wrong  number  of 
parameter" 

For example, CURRent 5.0,6. 

160 

"Unmatched 
quotation mark" 

For example, CALibrate:SECure 0,"6831'. 

165 

"Unmatched bracket" 

For example, CURRent (5. 

170 

"Invalid command" 

The  command  is  invalid  or  the  format  is 
wrong,  including  the  spelling  problems. 
For example, CUR 5.0. 

180 

"No entry in list" 

The  command  needs  three  parameters 
while  the  program  only  processes  two 
commands practically. 

191 

"Too many char" 

The sending data length is over 256 while 

using serial port or USB. 

-200 

"Execution error" 

The command execution is invalid due to 

wrong command status or wrong setting. 
For  example,  sending  *TRG  command 
must be under the BUS trigger method. 
If the trigger method is MANUAL, *TRG 
command is invalid. 

-310 

"System error" 

Undetermined error code. 

-350 

"Too many errors" 

Too much error in the stack. 

-410 

"Query 
INTERRUPTED" 

While  using  USB  communication,  no 
parameters  is  returned  after  sending 
query  command,  the  instrument  will 
prompt  an  error.  Some  status  or 
parameter  setting  command  need  to  be 
followed  by  a  query  command  i.e.  you 
need  to  send  the  query  command  after 
sending  this  setting  command,  if  you 
send  another  setting  command  after 
sending  this  setting  command,  the 
instrument will prompt an error. 

-430 

"Query 
DEADLOCKED" 

While  using  USB  communication,  if  you 
send  query  command  when  the  query 
region  is  full,  the  instrument  will  prompt 
an error. 

"No error" 

No error. 

"Mainframe 
Initialization Lost" 

During  the  start-up  program  initialization, 
if  the  instrument  can’t  read  the  user 
setting 

parameters 

and 

data  from 

EEPROM,  the  system  will  prompt  an 

Summary of Contents for IT6800A Series

Page 1: ...ower Supply Series IT6800A B Programming Guide Model IT6831A IT6832A IT6833A IT6835A IT6861A IT6862A IT6863A IT6872A IT6873A IT6874A IT6832B IT6833B IT6835B IT6861B IT6862B IT6863B IT6872B IT6873B IT6...

Page 2: ...terials in this document that conflict with these terms the warranty terms in the separate agreement shall prevail Technology Licenses The hardware and or software described herein are furnished under...

Page 3: ...es and other taxes shall be on the account of customer Limitation of Warranty This Warranty will be rendered invalid in case of the following Damage caused by circuit installed by customer or using cu...

Page 4: ...ll loading wires should be capable of bearing maximum short circuit current of power supply without overheating If there are multiple electronic loads each pair of the power cord must be capable of be...

Page 5: ...ffixed refers to the year when the design was approved The instrument complies with the WEEE Directive 2002 96 EC marking requirement This affixed product label indicates that you must not discard the...

Page 6: ...Ltd iv Product Category With reference to the equipment classifications described in the Annex I of the WEEE Directive this instrument is classified as a Monitoring and Control Instrument To return t...

Page 7: ...0 4 2 2008 EN 61000 4 2 2009 IEC 61000 4 3 2006 A1 2007 A2 2010 EN 61000 4 3 2006 A1 2008 A2 2010 IEC 61000 4 4 2004 A1 2010 EN 61000 4 4 2004 A1 2010 IEC 61000 4 5 2005 EN 61000 4 5 2006 IEC 61000 4...

Page 8: ...2 Chapter3 SCPI Essential Order 15 SYSTem ERRor 15 SYSTem VERSion 15 SYSTem REMote 15 SYSTem LOCal 16 SYSTem RWLock 16 SYSTem BEEPer 16 Chapter4 Trigger Commands 17 TRIGger 17 TRIGger SOURce 17 Chapte...

Page 9: ...tion SECure STATe 32 CALibration INITital 32 CALibration SAVe 32 CALibration VOLTage LEVel 32 CALibration VOLTage DATA numeric value 33 CALibration CURRent LEVel 33 CALibration CURRent DATA numeric va...

Page 10: ...t subsystem is listed below OUTPut SYNC OFF 0 ON 1 SYNC MODE NORMal CARRier POLarity NORMal INVerted OUTPut is the root class keyword SYNC is the second keyword MODE and POLarity are the third keyword...

Page 11: ...lev 3 the head of the second command curr is deleted and the instrument explains the second command as CURR PROT STAT OFF If curr is explicitly included in the second command it is semantically wrong...

Page 12: ...m and short form combination Note that each command word must be in long form or short form and not something in between For example SYSTe PRESe is illegal and will generate an error The command will...

Page 13: ...s completely spelled out such as VOLTAGE STATUS and DELAY In the short form the header has only the first three or four letters such as VOLT STAT and DEL Query Indicator Following a header with a ques...

Page 14: ...urns string parameters enclosed in double quotes Response messages A response message is the message sent by the instrument to the computer in response to a query command Sending a response message Af...

Page 15: ...ptions with given command strings are included in the brace The brace is not sent along with command strings Vertical stripes separate several parameter options with given command strings For example...

Page 16: ...by sending the following command TRIG COUN You may also query the allowable minimum or maximum count as follows TRIG COUN MIN TRIG COUN MAX Comma If a command requires several Parameter then a comma...

Page 17: ...meter to acceptable values The following commands require numerical Parameter of frequency value SOURce 1 2 FREQuency CENTer Frequency MINimum MAXimum NR1 There is an implicit decimal point in the las...

Page 18: ...1 6 1 RS232 interface There is a COM port DB9 connector at the rear of the power supply when connect to computer you need to connect a cable with COM port on both side To active connection you need t...

Page 19: ...t bit and a stop bit these values are fixed 2 As described before in RS 232 connector you must use the correct interface cable or adapter Note that even if the cable has the right plug the internal wi...

Page 20: ...rface as USB in the system menu the detailed operation is 1 Press Shift to enter the menu 2 Press to select SYST SET and then press P MEM KEEP is displayed then press to select COMM RS232 press 3 Pres...

Page 21: ...d Summary Register Enable Register not used not used not used not used not used not used STAT QUES STAT QUES ENAB value QUES 3 STAT QUES ENAB MAV 4 ESB 5 RQS 6 OPER 7 Event Register Enable Register Se...

Page 22: ...n chapter 5 5 CME 32 Command Error A command syntax error occurred see error numbers 101 through 178 in chapter 5 6 Not Used 0 Always set to 0 7 PON 128 Power On Power has been turned off and on since...

Page 23: ...ntly be changed Bit Decimal Value Definition 0 2 Not used 0 Always set to 0 3 QUES 8 One or more bits are set in the questionable status register bits must be enabled in the enable register 4 MAV 16 D...

Page 24: ...in the queue will be replaced by 350 Meaning too many mistakes If the error messages in the error message queue are not read out the error following will not be saved to the queue If there is no erro...

Page 25: ...ax SYSTem LOCal Parameter None SYSTem RWLock This order is used to set the power supply to remote control mode through the RS232 interface and LOCAL is not available Executing this order will set the...

Page 26: ...ax TRIGger IMMediate Parameters none Related orders TRG TRIG SOUR TRIGger SOURce This order is used to choose the source of trigger signal Power supply can receive the signal from front panel i e by p...

Page 27: ...is 0V 1mA Order syntax OUTPut STATe bool Parameters 0 1 OFF ON Query syntax OUTPut RST value 0 OUTPut TIMer This order is used to set the state of output timer Order syntax OUTPut TIMer STATe bool Pa...

Page 28: ...Output Commands Copyright Itech Electronic Co Ltd 19 Parameters NRf...

Page 29: ...rders to increase or decrease current the stepping value should be set with order CURR STEP Oder syntax SOURce CURRent LEVel IMMediate AMPLitude NRf Parameters MIN to MAX Unit A RST value MAX Query sy...

Page 30: ...Current Control Commands Copyright Itech Electronic Co Ltd 21 Parameters MIN to MAX RST value the stepping resolution of current of this unit Query order CURRent STEP Return parameters NR2...

Page 31: ...s to increase or decrease voltage the stepping value should be set with order VOLTage STEP Order syntax SOURce VOLTage LEVel IMMediate AMPLitude NRf Parameters MIN to MAX Unit V RST value MIN Query sy...

Page 32: ...it of OVP then we can assume that internal parts has been shorted Query state register OV bit has been set When power supply is protected then we can through order VOLT PROT CLE to clear the OVP state...

Page 33: ...der the output value will be restored to the previous output and the upper limit of OVP will be kept Of course before you sending this order you should enable the output voltage lower than the upper l...

Page 34: ...Voltage Control Commands Copyright Itech Electronic Co Ltd 25 Order syntax SOURce VOLTage RANGe high low Parameters high low...

Page 35: ...age and current according to the current setting of this order Of course the precondition is the setting value is within the set range If not then a execution error will occur You can also use DEF MIN...

Page 36: ...icate with PC and sending the current data to PC This order will not affect our unit s setting or trigger the measurement operation It only needs the nearest reading it can get The returned readings w...

Page 37: ...Sure POWer This order is used to measure the current output power value Order syntax MEASure SCALar POWer DC Return parameters NR2 FETCh POWer This order can read the latest power value from the sampl...

Page 38: ...Tage This order is used to set the steps of list file and voltage of current step Order syntax SOURce LIST VOLTage NRf Parameters MIN to MAX Unit V Example LIST VOLT 1 3V set the first step s voltage...

Page 39: ...URR 1 inquire the first step s current Return parameters NR2 LIST TIMEr This order is used to set the steps of list file and delay time of current step Order syntax SOURce LIST TIMEr Parameters 0 1 to...

Page 40: ...order can recall the list file saved before from specified memory region Sending query orders can inquire which group file you have recalled Order syntax SOURce LIST LOAD IMMediate NR1 Parameters 0 to...

Page 41: ...ument model CAL SEC OFF Query syntax CALibration SECure STATe CALibration INITital This order is used to reset the calibration parameters to default setting Order syntax CALibration INITital Parameter...

Page 42: ...2 must be calibrated orderly Order syntax CALibration CURRent LEVel point Parameters P1 P2 CALibration CURRent DATA numeric value Return actual output current value to calibration point Order syntax C...

Page 43: ...Electronic Co Ltd 34 time number of times and so on Example CAL STR 2005 1 9 20 12 CALibration STRing Check original calibration information Query syntax CALibration STRing Return parameters calibrati...

Page 44: ...egister Quest condition register Status byte register Order syntax CLS Parameter None ESE This order can set the parameter of standard event enable register Setting parameter can determine which bit v...

Page 45: ...used 0 OPC Operation complete 1 ESE This order can read the value of standard event status register Quest syntax ESE Parameter None Return parameter NR1 ESR This order can read the value of standard...

Page 46: ...Return parameter AARD Example ITECH 6800A 00000000000004 V1 01 V1 00 OPC When all orders before this order are executed OPC is 1 for the standard event status register Sending query order will return...

Page 47: ...RST Parameter None SRE enabled value This order can set the parameter of status bit enable register When query status bit enable register the power will return a decimal number this number is the bina...

Page 48: ...te register is cleared while the status bit will not be cleared when system statistics implemented Quest syntax STB Parameter None Return parameter NR1 Reference order CLS ESE ESR TRG When the trigger...

Page 49: ...EStionable EVENt This order can read the parameter from quest event register The power supply will return a decimal number which is corresponding to the binary weighted sum of each bit of the register...

Page 50: ...r is OFF or uncertain If 1 is returned then the status is CV mode if 2 is returned then the status is CC If 3 is returned then there is error occurred Quest syntax STATus QUEStionable CONDition Parame...

Page 51: ...port or USB 200 Execution error The command execution is invalid due to wrong command status or wrong setting For example sending TRG command must be under the BUS trigger method If the trigger method...

Page 52: ...ou can t enter the calibration mode the system will prompt an error For example for model IT6831A the correct calibration command is CALibrate SECure 0 6831 if you send CALibrate SECure 0 6811 the sys...

Page 53: ...roducts If you have any doubt about this product please contact us as follow 1 Please refer to the CD ROM of related user s manual in package 2 Visit ITECH website www itechate com 3 Select the most c...

Reviews: