IT-E151AUser Manual
Copyright © ITECH Electronics Co., Ltd.
iii
IEC 61000-4-3:2002. Radiofrequency electromagnetic field immunity
4
IEC 61000-4-4:2004. Electrical fast transient/burst immunity.
IEC 61000-4-5:2001. Power line surge immunity.
IEC 61000-4-6:2003. Immunity to conducted disturbances induced by
radiofrequency
5
IEC 61000-4-11:2004. voltage dips, short interruptions and voltage
variations immunity
6
EN 61000-3-2:2006: Harmonic emission on AC power line.
EN 61000-3-3:1995: Voltage change, fluctuation and flicker
1 The product is intended for use in non-residential/non-domestic environments. Use of
the product in residential/domestic environments may cause electromagnetic
interference.
2 Connection of the instrument to a test object may produce radiations beyond the
specified limit.
3. Use high-performance shielded interface cable to ensure conformity with the EMC
standards listed above.
4 Trace bloom not exceeding 4 divisions pk-to-pk may be induced under the test
conditions specified in IEC 61000-4-3.
5 Trace bloom not exceeding 1 divisions pk-to-pk may be induced under the test
conditions specified in IEC 61000-4-6.
6 This specification C is applied to 70% voltage drop in 25 cycles and 0% voltage
interruption in 250 cycles subject to IEC 61000-4-11.