
Europark Oost 24
Tel : ++32 3 766 02 02
E-mail : [email protected]
9100 Sint-Niklaas – Belgium
Fax : ++32 3 766 10 67
12
Blends such as AZ-50, HP62, MP39 etc…
Detect Ethylene Oxide gas leaks in hospital sterilizing equipment (it will detect the
halogenated carrier gas)
Detect SF-6 in high voltage circuit breakers
Detect most gases that contain Chlorine, Fluorine and Bromine (halogen gases)
Detect cleaning agents used in dry cleaning applications such as perchloroethylene
Detect Halon gases in fire extinguishing systems
Maintenance
Proper maintenance of you Leak Detector is very important. Carefully following the
instructions, outlined below, will reduce performance problems and increase the live
expectancy of the unit.
WARNING
: TURN UNIT OFF BEFORE REPLACING THE SENSING TIP. FAILURE TO DO
SO MAY RESULT IN A MILD ELECTRICAL SHOCK!
Keep the sensing tip clean: Prevent dust, moisture and grease build-up by utilizing the
provided tip protector. Never use the unit without the protector in place.
Before using the unit always inspect the tip and protector to see that they are free of
dirt and/or grease. To clean:
1.
Remove protector by grasping and pulling off tip.
2.
Clean protector with shop towel and/or compressed air.
3.
If the tip itself is dirty it can be cleaned by immersing in a mild solvent, such as
alcohol, for a few seconds, and then using compressed air and/or a shop towel to clean.
NOTE: Never use solvents such as gasoline, turpentine, mineral spirits, etc… as these will
leave a detectable residue and desensitize your unit.
Sensing tip replacement
The tip will eventually wear out and require replacement. It is difficult to predict exactly
when this will occur since tip longevity is directly related to the conditions and frequency
of use. The tip should be replaced whenever the alarm sounds or becomes erratic, in a
clean, pure, air environment.
To replace the tip:
1.
Make sure the unit is OFF.
2.
Remove the old tip by unscrewing counter-clockwise.
3.
Use the supplied replacement tip, located in the carrying case. Replace by screwing on
clockwise.