background image

 

 

 

 

 
 

 

 

Europark Oost 24 

Tel  : ++32 3 766 02 02 

 

E-mail : [email protected] 

 

9100 Sint-Niklaas – Belgium 

Fax  : ++32 3 766 10 67 

 

 

 

10 

 

Operating tips 

 

The following section includes several general operating tips, and the SAEJ1628 
recommended procedure for leak detection. 

 

1.

 

Adjust the sensitivity up, only when a leak cannot be found. Adjust the sensitivity down 
only when resetting the unit does not allow you to „home in‟ on the leak. 

2.

 

In areas that are heavily contaminated with gas, the unit may be reset to block out 
ambient concentrations of gas. The probe should not be moved while the unit is being 
reset. The unit can be reset as many times as needed. 

3.

 

In windy areas, even a large leak can be difficult to find. Under these conditions, it is 
best to shield the potential leak area. 

4.

 

Be aware that the detector may alarm if the sensing tip comes in contact with 
moisture and/or solvents. Therefore, avoid contact with these when leak checking. 

 

SAE J1628 Recommended Procedure 

 
NOTE: On Automotive A/C Systems leak test with the engine not in operation. 
 
1.

 

The air conditioning or refrigeration system should be charged with sufficient 
refrigerant to have a gauge pressure of at least 340 kPa (50 psi) when not in 
operation. At temperatures below 15°C (59 °F), leaks may not be measurable, since 
this pressure may not be reached. 

2.

 

Take care not to contaminate the detector probe tip if the part being tested is 
contaminated. If the part is particularly dirty, or condensate (moisture) is present, it 
should be wiped off with a dry shop towel or blown off with shop air. No cleaners or 
solvents should be used, since the detector may be sensitive to their ingredients. 

3.

 

Visually trace the entire refrigerant system, and look for signs of air conditioning 
lubricant leakage, damage and corrosion on all lines, hoses, and components. Each 
questionable area should be carefully checked with the detector probe, as well as all 
fittings, hose to line couplings, refrigerant controls, service ports with caps in place, 
brazed or welded areas, and areas around attachment points and hold-downs on lines 
and components. 

4.

 

Always follow the refrigerant system around in a continuous path so that no areas of 
potential leaks are missed. If a leak is found, always continue to test the remainder 
of the system. 

5.

 

At each area checked, the probe should be moved around the location, at a rate no 
more than 25 to 50 mm/second (1-2 in/second), and no more than 5 mm (1/4 in) from 
the surface, completely around the position. Slower and closer movement of the 
probe greatly improves the likelihood of finding a leak (see fig.7). Any increase in 
beep rate is indicative of a leak. 
 

Summary of Contents for ITE-XTRA

Page 1: ...Europark Oost 24 Tel 32 3 766 02 02 E mail sales ite tools com 9100 Sint Niklaas Belgium Fax 32 3 766 10 67 1 ITE XTRA ...

Page 2: ...formation 3 Features 3 Parts Controls 4 Getting started 4 Installing batteries 4 Operating features 5 Battery indicator 5 Automatic reset 6 Sensitivity adjustment 6 Alarm indications 7 Operating instructions 7 Operating tips 8 Applications 9 Maintenance 9 Replacement parts 10 Specifications 11 Warranty 11 Español 12 Français 20 Deutsch 29 ...

Page 3: ...een added to enhance the usability of the ITE XTRA Seven levels of sensitivity provide an increase of 64 times from level 1 to level 7 Unique Tri color led s show a progressive and wide ranging leak size indications communicate the sensitivity level and provide a true voltage indication of battery power level A tactile keypad controls all functions of operation A revolutionary new case design give...

Page 4: ...eference leak source Parts controls 1 Sensing tip 6 Audio mute 2 Tip protector 7 Sensitivity up 3 Power ON OFF 8 Sensitivity down 4 Battery test 9 LED leak indicators 5 Reset button 10 Flexible probe Getting started Installing batteries Remove the battery compartment door located on the bottom of the instrument by sliding up as shown below Install batteries positive polarity outwards towards batte...

Page 5: ...on The constant power indicator allows the user to see the battery level at all times The LED will remain on whenever the unit is powered on It may appear as one of three colors See Fig 3 GREEN Battery voltage is normal sufficient for proper operation ORANGE Battery voltage is approaching the lower threshold for operation replace as soon as possible RED Battery voltage is below acceptable operatin...

Page 6: ...unit automatically sets itself to ignore the level of refrigerant present at the tip Only a level or concentration greater than this will cause an alarm CAUTION Be aware that this feature will cause the unit to ignore any refrigerant present at turn on In other words with the unit off if you place the tip up to a known leak and switch the unit on no leak will be indicated RESET FEATURE Pressing th...

Page 7: ...display will show the LED s red The number of LED s lit indicates the level See Fig 5 Level one lowest sensitivity is shown by the leftmost LED Counting from left levels 2 through 7 are indicated by the corresponding number of red LED s i e level 7 is shown by all LED s lit 2 Pressing the SENSITIVITY or SENSITIVITY key will change the sensitivity The keys can be pressed intermittently to change le...

Page 8: ...e Orange one at a time Operating Instructions Operation 1 Switch the unit on by pressing the ON OFF key The display will illuminate with the reset indication Left LED green all others Orange for 2 seconds 2 Verify the battery level by observing the constant power indicator see above 3 Upon turn on the unit is set to sensitivity level 5 A rapid but steady beep rate will be heard If desired the sens...

Page 9: ...32 3 766 02 02 E mail sales ite tools com 9100 Sint Niklaas Belgium Fax 32 3 766 10 67 9 6 If a full alarm occurs before the leak is pinpointed press the RESET key to reset the circuit to a zero reference as described above ...

Page 10: ...mperatures below 15 C 59 F leaks may not be measurable since this pressure may not be reached 2 Take care not to contaminate the detector probe tip if the part being tested is contaminated If the part is particularly dirty or condensate moisture is present it should be wiped off with a dry shop towel or blown off with shop air No cleaners or solvents should be used since the detector may be sensit...

Page 11: ... minimum shutting it off then waiting for the refrigerant to accumulate in the case for 10 minutes After such time insert the leak detector probe into the blower resistor block or condensate drain hole if no water is present or into the closest opening in the heating ventilation air conditioning case to the evaporator such as the heater duct or a vent duct If the detector alarms a leak apparently ...

Page 12: ...tip protector Never use the unit without the protector in place Before using the unit always inspect the tip and protector to see that they are free of dirt and or grease To clean 1 Remove protector by grasping and pulling off tip 2 Clean protector with shop towel and or compressed air 3 If the tip itself is dirty it can be cleaned by immersing in a mild solvent such as alcohol for a few seconds a...

Page 13: ...pped with one Carrying Case one Owner s Manual 2 C cell batteries and one replacement Sensing Tip and Protector To purchase replacement parts for your leak detector please contact your local distributor To ensure that you obtain the correct parts it is best to reference the part number when placing your order Replacement Parts ITE XTRA 2 Maintenance Kit 3 Sensing Tips 3 Tip Protectors ...

Page 14: ...ured to provide unlimited service Should the unit be inoperative after performing the recommended maintenance a no charge repair or replacement will be made to the original purchaser if the claim is made within three years from the date of purchase This warranty applies to all repairable instruments that have not been tampered with or damaged through improper use This warranty does not cover batte...

Reviews: