background image

4

POWER-UP THE MFM300   

Press the POWER button until the display shows the Cooper logo, the remaining battery life and 

the installed firmware. Press the LIGHT button to illuminate the screen in low light conditions.

POWER-DOWN THE MFM300

Press the POWER button until the display goes blank. 

DISPLAY MODES

Normal mode: Displays the measured values of all the probes currently connected. 

Superheat mode: Displays superheat or sub-cooling values. Requires a temperature probe be 

installed in either T1 or T2 (or both) and optionally a pressure probe in U1 or U2 (or both). 

Press the SH button to enter Superheat Mode. Press the SH button again, or CANCEL, to return 

to Normal Mode. See the section on measuring system Superheat / Sub-cooling for further 

information.

Psychrometrics mode: Displays psychrometric values. Requires a relative humidity probe to be 

installed in either U1 or U2 (or both). Press the PSYCH button to enter the Psychrometric Mode

Press the PSYCH button again, or CANCEL, to return to Normal Mode.

The MFM300 will attempt to operate in the last mode it was in when powered off. If it can not, the 

Normal Mode is the default. If no probes are connected, the “No Probes” message will appear on 

the display.

DISPLAY MIN, MAX AND AVG VALUES

When in Normal Mode press:

 MIN 

button to display the lowest readings sensed by each probe.

 MAX button to display the highest readings sensed by each probe.

 AVG button to display the average readings sensed by each probe.

The active mode, MIN, MAX or AVG, will be indicated near the bottom of the display. Pressing 

the same button again, or CANCEL, turns off the selected mode.  Disconnecting a probe will 

clear that probe’s MIN, MAX, and AVG memory, but all probes still connected will continue to 

be updated. All MIN, MAX and AVG memory is lost when the MFM300 is powered down.

MEASURE TEMPERATURE

Install one, two, or three Cooper 10K thermistor probes into any of the three temperature jacks: 

T1, T2 or T3. The MFM300 will sense the probe and display the temperature measurement with 

the appropriate label.

MEASURE TEMPERATURE DIFFERENCE (DELTA T)

With two probes connected, press the DELTA button until “Delta T” appears on the display. The 

absolute temperature difference appears to the right of the horizontal bars that point to the two 

temperature inputs being measured. Press the DELTA button again, or CANCEL, to turn off the 

“Delta T” display.  If three temperature probes are connected, press the DELTA button again to 

display the next “Delta T”. With three temperature probes connected, the “Delta T” sequence is 

as follows:

 

Summary of Contents for Cooper MFM300

Page 1: ...MFM300 Multi Function Instrument User Guide and Instrument Warranty...

Page 2: ...te Superheat Sub Cooling 5 Measure Dynamic System Superheat Sub Cooling 5 6 Measure Relative Humidity and Dry Bulb Temperature 6 Display Psychometrics 6 Measure Air Flow 6 Air Balancing 6 Navigate the...

Page 3: ...JACK U1 UNIVERSAL JACK U2 POWER TEMPERATURE JACK T2 TEMPERATURE JACK T1 DOWN ARROW PSYCHROMETRICS PSYCH TEMPERATURE JACK T3 SHIFT SUPERHEAT SUB COOLING SH UP ARROW ENTER REFRIGERANT REF MENU AVERAGE...

Page 4: ...ssage will appear on the display DISPLAY MIN MAX AND AVG VALUES When in Normal Mode press MIN button to display the lowest readings sensed by each probe MAX button to display the highest readings sens...

Page 5: ...memory There is room in the memory for up to five more refrigerants New refrigerant P T data can be downloaded from the Cooper Atkins website www cooper atkins com mfm asp using the MFM300 Toolbox an...

Page 6: ...layed you must enter the shape and size of the duct opening where airflow is being measured To do this press the MENU button while viewing CFM Use the on screen directions to enter the duct shape and...

Page 7: ...Place a temperature probe connected to T1 into an environment with a known temperature and adjust the reading to calibrate and match Zero out probe P1 Zero out probe P2 When a pressure transducer is...

Page 8: ...ace it where temperatures are to be measured Download data Select this menu option to store the Data Logger contents in the MFM300 and view at a later time View data Select this menu option to view th...

Page 9: ...rature Probes Temperature Measurement Range 58 to 302 F 50 to 150 C Temperature Accuracy 0 3F 0 2C or 0 5 of Reading whichever is greater Temperature Display Resolution 0 1 Degree Relative Humidity RH...

Page 10: ...es notice that any statements or representations made by it will not create a warranty that this product is fit for any particular purpose Such statements representations or descriptions are informati...

Reviews: