TTI-10 Iss.02 – 07/11 

3) Calibration by Comparison to a Standard Thermomter / oP2 

 

Use the up and down keys [



] to select 

oP2

. Push [ENTER/MENU] to confirm. 

On  the  bottom  of  the  display  appears 

1  P

.  For  changing  between  a  1-Point 

1  P

,  2-Point 

2  P

  or  3-Point 

3  P

  - 

calibration use the up and down keys [



]. 

 

 

oP2 

 

 

 

 

 

 

1 P 

 

2 P 

  3P   

 

Example of a 1-Point calibration: 

 

Push [ENTER/MENU] to confirm. On the display appears 

CALC

. After the displayed measuring value is stabile push 

[ENTER/MENU]. On the first display line you can see the “frozen” measurement value. On the second line as a 

default you can see 

-100.000

. Now you have to Enter (instead of -100.000) the correct measurement value from 

your reference: 

 

By using the up key [

] you are able to move the decimal point to setup the number of decimal places. Push 

[ENTER/MENU] to confirm. 

 

Now  the  algebraic  sign  is  blinking

„-„

.  Use  the  up  key  [

]  to  toggle  for  positive  or  negative  number.  Change  the 

number using the up and down keys [



]. 

 

Note: 

Up key 

 is changing the blinking segment 

Down key

 is jumping to the next segment 

Push [ENTER/MENU] to confirm, revert back to the measuring mode. 

 

Important: 

A break of the physical calibration cannot be done by the [ESC]-button. 

 

Display-indication with active calibration code (OP2): 

 

The 

CAL

-segment and the small 

2

 indicates to the user that 

oP2

 is activated. 

 

4) Calibration according to Coefficients of EN60751 (R0,ABC) 

 

By using Option 3 you are able to activate coefficients according to EN60751 (R0,A,B,C). The coefficients have to be 

calculated using special software on a PC (e. g. P7_CALC). Before you are able to activate this option you have to 

transmit the coefficients from the PC to the instrument.  

 

Use the up and down keys [



] to select 

oP3

. Push [ENTER/MENU] to confirm. 

 

Now the calibration option 3 is activated. Push [ESC] to revert back to the measuring mode. 

 

Note: 

After confirming 

oP3 

by pushing [ENTER/MENU] the function 

oP3

 (calibration by code) is activated, even 

though you leave the menu by pressing [ESC]. 

 

Display-indication with active calibration code (OP3): 

 

The 

CAL

-segment and the small 

3

 indicates to the user that 

oP3

 is activated. 

 

5)  oP4 

Reserved for future use, not currently implemented 

Summary of Contents for TTI-10

Page 1: ...4 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate Equally because of the progra...

Page 2: ...C F and Ohms 6 Temperature Difference Mode Line2 6 Calibration function CAL 6 Channel Activation Chnl 9 Memory Setup Lo6 9 Recalling the memory data HOLD MAX MIN AVE 10 Measuring rate Normal Mode FAS...

Page 3: ...PARTS ARE NOT COVERED BY THIS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prol...

Page 4: ...ERSONNEL FAMILIAR WITH GOOD MEASUREMENT PRACTICES IT IS EXPECTED THAT PERSONNEL USING THIS EQUIPMENT WILL BE COMPETENT WITH THE MANAGEMENT OF APPARATUS WHICH MAY BE POWERED OR UNDER EXTREMES OF TEMPER...

Page 5: ...vice personnel Contact Isothermal Technology or one of our appointed distributors for details of approved service outlets Operation Before switching on the instrument connect the probe or probes to th...

Page 6: ...o adjust the requested selection Push ENTER MENU to confirm Push ESC to revert back to the measuring mode MENU Unit Lin2 CAL Chnl Log The calibration is hidden when the instrument has been calibrated...

Page 7: ...uested calibration option Push ENTER MENU to confirm CAL oFF oP1 oP2 oP3 oP4 1 Standard calibration according IEC 60751 oFF No Corrections are applied Use the up and down keys to select oFF Push ENTER...

Page 8: ...ing the blinking segment Down key is jumping to the next segment Push ENTER MENU to confirm revert back to the measuring mode Important A break of the physical calibration cannot be done by the ESC bu...

Page 9: ...p and down keys again to start on or Stop off the logger Push ENTER MENU to confirm Use the up and down keys again to select between automatic storage Auto or manually operated storage SPot Push ENTER...

Page 10: ...Mode reduced resolution 0 01 from 199 99 to 199 99 Filter Mode high resolution 0 001 from 199 999 to 199 999 Press FAST key to change the measuring rate By using the Filter Mode the instrument is per...

Page 11: ...llows app 1 hour of further operation The battery symbol indicates according to the battery status between 1 to 3 segments We recommend the use of high capacity alkaline batteries such as the Duracell...

Page 12: ...o 199 999 C remaining range 0 01 C Accuracy Instrument Only 0 01 C from 80 C to 199 999 0 02 C 0 0015 C reading from 200 C to 660 C Logging Record Average Min and Max over 4000 measurements Measuring...

Reviews: