background image

                                   

Page 11 of 12 

TTI-10 Iss.02 – 07/11 

TARE-Function (ZERO-Mode F1) 

 

The Instrument has a special Tare/Zero-button. Pressing the [F1] key once, the instrument will subtract the last 

measured value from the actual measured value. So if the measurement does not change zero appears on the 

instrument’s display. If you press the [F1] key once again the instrument move to the Normal-Mode. 

 

E.g. Display with activated Zero-Mode: 

The arrow on the left indicates that the instrument is working in the Zero-Mode [F1]. 

Note: 

After switching off the instrument, this function is automatically deactivated. 

Power supply 

 

For the power supply of the instrument a 9 Volt dry battery is used. To exchange the battery switch of the instrument 

and open the rear battery cover. Remove the battery from the instrument and replace with a new battery.  

 

For operation without a battery the TTI-10 can be powered from via the supplied USB Cable. 

 

The „BAT“symbol in the display indicates that the battery needs to be exchanged. After displaying the “BAT” symbol, 

the instrument allows app. 1 hour of further operation. 

 

The battery symbol indicates according to the battery status between 1 to 3 segments. 

 

We recommend the use of high capacity alkaline batteries such as the Duracell Procell or rechargeable capacity NiMH 
Cells of 250mAh. 

 

Note:

 For protection of our environment please don’t put the battery into general household waste, but use a local 

authority approved disposal method. 

Error Codes 

 

Error   

Meaning 

Open   

no probe or wrong probe is connected 

 

Use with Cal Notepad 

 

The TTI-10 can be used with Cal Notepad software. 

First download the Driver and the Cal Notepad 

Software from 

http://isotech.info/TTI-10

 Unzip the 

driver and making a note of the file location. 

 

Connect the TTI-10 to the PC with the USB lead and 

when prompted point the Device manager to the 

location of the driver files. The PC will then assign a 

Com Port number to the instrument, this number can 
be found from the Device Manager.  

 

Next install Cal Notepad, start the application and in the 

instrument tab set the Com Port to the value as above. 

Click start and the TTI1-10 can be monitored and the 

data logged as required. The TTI-10 will log in whatever 

units have been selected from the Units menu; see 

Selecting Display Units on page 6.  

Summary of Contents for TTI-10

Page 1: ...4 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate Equally because of the progra...

Page 2: ...C F and Ohms 6 Temperature Difference Mode Line2 6 Calibration function CAL 6 Channel Activation Chnl 9 Memory Setup Lo6 9 Recalling the memory data HOLD MAX MIN AVE 10 Measuring rate Normal Mode FAS...

Page 3: ...PARTS ARE NOT COVERED BY THIS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prol...

Page 4: ...ERSONNEL FAMILIAR WITH GOOD MEASUREMENT PRACTICES IT IS EXPECTED THAT PERSONNEL USING THIS EQUIPMENT WILL BE COMPETENT WITH THE MANAGEMENT OF APPARATUS WHICH MAY BE POWERED OR UNDER EXTREMES OF TEMPER...

Page 5: ...vice personnel Contact Isothermal Technology or one of our appointed distributors for details of approved service outlets Operation Before switching on the instrument connect the probe or probes to th...

Page 6: ...o adjust the requested selection Push ENTER MENU to confirm Push ESC to revert back to the measuring mode MENU Unit Lin2 CAL Chnl Log The calibration is hidden when the instrument has been calibrated...

Page 7: ...uested calibration option Push ENTER MENU to confirm CAL oFF oP1 oP2 oP3 oP4 1 Standard calibration according IEC 60751 oFF No Corrections are applied Use the up and down keys to select oFF Push ENTER...

Page 8: ...ing the blinking segment Down key is jumping to the next segment Push ENTER MENU to confirm revert back to the measuring mode Important A break of the physical calibration cannot be done by the ESC bu...

Page 9: ...p and down keys again to start on or Stop off the logger Push ENTER MENU to confirm Use the up and down keys again to select between automatic storage Auto or manually operated storage SPot Push ENTER...

Page 10: ...Mode reduced resolution 0 01 from 199 99 to 199 99 Filter Mode high resolution 0 001 from 199 999 to 199 999 Press FAST key to change the measuring rate By using the Filter Mode the instrument is per...

Page 11: ...llows app 1 hour of further operation The battery symbol indicates according to the battery status between 1 to 3 segments We recommend the use of high capacity alkaline batteries such as the Duracell...

Page 12: ...o 199 999 C remaining range 0 01 C Accuracy Instrument Only 0 01 C from 80 C to 199 999 0 02 C 0 0015 C reading from 200 C to 660 C Logging Record Average Min and Max over 4000 measurements Measuring...

Reviews: