7.2.3 Calibrating the 100
Ω
Internal Reference Resistor
Repeat the procedure for calibrating the 400
Ω
internal reference (see section
7.2.2), but configure the microK to measure a Default PRT at 1mA against the
internal 100
Ω
reference on the 125
Ω
range with channel 1. The command to
adjust the value assigned to the 100
Ω
internal reference is:
CAL:REF204
R100
Where
R100
is the value calculated for the 100
Ω
internal reference.
7.2.4 Calibrating the 25
Ω
Internal Reference Resistor
Repeat the procedure for calibrating the 400
Ω
internal reference (see section
7.2.2), but configure the microK to measure a Default PRT at 1mA against the
internal 25
Ω
reference on the 125
Ω
range with channel 1. The command to
adjust the value assigned to the 25
Ω
internal reference is:
CAL:REF203 R25
Where
R25
is the value calculated for the 25
Ω
internal reference.
7.2.5 Calibrating the 10
Ω
Internal Reference Resistor
Connect a stable external resistance standard with a nominal value of 10
Ω
to
channel 1. Ideally, this should be a Wilkins resistor with a recent (traceable)
calibration certificate maintained at a constant temperature in an oil bath. Repeat
the procedure for calibrating the 400
Ω
internal reference (see section 7.2.2), but
configure the microK to measure a Default PRT at 5mA against the internal 10
Ω
reference on the 25
Ω
range with channel 1. The command to adjust the value
assigned to the 10
Ω
internal reference is:
CAL:REF202
R10
Where
R10
is the value calculated for the 10
Ω
internal reference.
7.2.6 Calibrating the 1
Ω
Internal Reference Resistor
Connect a stable external resistance standard with a nominal value of 1
Ω
to
channel 1. Ideally, this should be a Wilkins resistor with a recent (traceable)