Page 27 of 41
Medusa 510/511 Series Iss.11 – 07/12
Testing Thermostats
The Site model can be used the Isotech Cal Notepad software for the testing of thermostats and other thermal
switches with volt free contacts. Cal Notepad can capture the temperature at which a switch opens or closes. It can
also perform a hysteresis test. Refer to the Cal Notepad manual for details. Ensure only voltage free contacts are
connected. Do not allow any voltage signal to be connected - doing so may damage the equipment and present a
safety hazard.