Medusa 510/511 Series Iss.11 – 07/12
CALIBRATION DATA EXAMPLE
A maximum of five points may be entered, shown as Pnt 1 to Pnt 5 for the temperature point and Ofs 1 to Ofs 5 for
the offset values.
The Pnt values must be entered in ascending order.
Set a Pnt to a value lower than the previous point to disable it.
The indicator would be programmed with the following data:
Pnt 1 100
Ofs 1 0.8
Pnt 2 300
Ofs 2 1.1
Pnt 3 500
Ofs 3 2.1
Pnt 4 -999
Ofs 4 0
Pnt 5 -999
Ofs 5 0