OPERATING INSTRUCTIONS 

INTRODUCTION 

 

You or your company has just paid a considerable sum of money to purchase this 96178. 

 

Although the materials to make the 96178 are very costly, it has also taken up to 1000 hours to prepare the 96178 for 

your use. 

 

It is not a product that can be assembled, tested and sold; it is produced, calibrated, aged and re-calibrated, until its 

characteristics are stable enough to meet the exacting needs of ITS 90. 

 

The weeks of work that go into its production make each 96178 more than another product.  Each 96178 already has 

a character and a history before it leaves us.  Please look after it.  Regard yourself as its custodian rather than its 

owner. 

 

1. 

Always keep the 96178 in its case when not in use. 

 

2. 

When in use, support the handle. 

 

3. 

Cool and store the 96178 in the same place as you normally use it. 

 

4. 

Each time before you use the 96178 clean off all traces of grease by using a chlorinated solvent. 

 

5. 

Even go so far as using gloves to handle the 96178, and keep the gloves clean. 

 
6. 

Quartz is glass.  It is a supercooled liquid.  At 800 C and above, your 96178 will bend and bow if you do not 

support it along its complete length.  At high temperatures I recommend that the 96178 be housed inside a 
close fitting recrystallised Alumina closed ended tube, which has been pre-fired to 1000 C, or better still a 

silicon carbide tube. 

 
7. 

Quartz is transparent in two senses of the word.  At temperatures above 700 C metallic vapours can pass 

through the quartz and attack the pure platinum sensing element.  Isotech have developed a product which 
can be attached to the 96178 to prevent this happening to a 960 Ioniser.  A Model 960 Ioniser is provided 

free with each 96178. 

 

8. 

If you have purchased the „Open‟ version of the 96178, open the valve at high temperatures and close it 

before bringing the 96178 down to room temperature.  This will ensure that no moisture builds up inside the 

thermometer. 

 

RETURNING YOUR THERMOMETER TO ISOTECH 

 

Due to the fragility of the 96178 we strongly recommend that this is not transported unless personally hand carried. 
 

In the event that the unit must be shipped by courier then the following instructions must be adhered to and 

reference to diagram on page 12. 

 

Summary of Contents for 96178

Page 1: ...PR9 9AG England Tel 44 0 1704 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate E...

Page 2: ...TO ISOTECH 4 Preparing the box 5 STABILITY EVALUATION 6 INTRODUCTION 6 METHOD 6 RESULTS 7 DISCUSSION OF RESULTS 7 CONCLUSION 8 FURTHER WORK 8 GLOSSARY OF METROLOGICAL TERMS 11 ACCURACY OF MEASUREMENT...

Page 3: ...GLASS PARTS ARE NOT COVERED BY THIS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will b...

Page 4: ...8 and keep the gloves clean 6 Quartz is glass It is a supercooled liquid At 800 C and above your 96178 will bend and bow if you do not support it along its complete length At high temperatures I recom...

Page 5: ...from moving seal the box lower it into the outer box and fill with chips then lift the inner box slightly so the polystyrene chips completely surround it Place the protective sheath if supplied on to...

Page 6: ...thermometers will change its characteristics if subjected to step changes of temperature The way in which the thermometer is temperature cycled is therefore very important In this series of tests the...

Page 7: ...ure even so the stability of the new thermometer is remarkable There is always an initial shift in the characteristics of a high temperature thermometer during its first cycle to the silver point In t...

Page 8: ...e Uncertainties vary depending on the temperature range of calibration Isotech s Models 670H and 96178 Standard Platinum Resistance Thermometers SPRT s can be calibrated using procedural techniques ai...

Page 9: ...Page 9 of 12 HTSPRT Model 96178 Iss 06 03 11...

Page 10: ...Page 10 of 12 HTSPRT Model 96178 Iss 06 03 11...

Page 11: ...2 Reproducibility may be expressed quantitatively in terms of the dispersion of the results UNCERTAINTY OF MEASUREMENT An estimate characterising the range of values within which the true value of a...

Page 12: ...Page 12 of 12 HTSPRT Model 96178 Iss 06 03 11 PACKING INSTRUCTIONS FOR THE RETURN OF YOUR 96178...

Reviews: