6
a.Set the rotary switch at desired range
b.Press the trigger to open the jaw and fully enclose all two
wires(single phase, two wires), three wires(three phases, three wires),
or four wires(three phases, four wires). Make sure the two half jaws
are properly closed.
c. Read the measured value from the LCD display.
3.1.3. Using the 50/60 and Wide Selector
50/60 Position
This clamp meter has very good frequency response due to the
electric property of the transformer jaws used. Therefore, the
measurement result contains not only the fundamental frequency of
50/60Hz but also the high frequencies and harmonics superimposed
on the fundamental frequency. To eliminate the effect of high
frequency noise, a low pass filter is designed to filter out high
frequency signal. To enable the filter, set the switch at the 50/60
position. The filter's cut-off frequency is set at 100Hz with an
attenuation characteristic of approx. 24dB/octave.
Wide Position
If the circuit under test is originated from a high frequency generating
device such as inverter, switching regulators, etc., then the switch
should be set at wide position to measure the signal which contains
the frequency from 40Hz - 1KHz.
To make sure the presence of high frequency signal, set the switch at
50/60 and wide position to see the difference. If the reading is very
different, it is certain that the high frequency signals or harmonics
present.
3.2. AC Load Current Measurement
Summary of Contents for ILCM 03A
Page 1: ...Instruction Manual ILCM 03A AC Leakage Current Tester FR EN IT DE ES...
Page 7: ...5 3 1 2 Out of Balance Leakage Current...
Page 11: ...9 turning on power...
Page 15: ...Manuel d utilisation ILCM 03A Testeur de Courant de Fuite FR EN IT DE ES...
Page 21: ...5 3 1 2 Courant de fuite d s quilibr...
Page 29: ...Manuale di istruzioni ILCM 03A Pinza Amperometrica per Correnti di Perdita FR EN IT DE ES...
Page 35: ...5 3 1 2 Tenuta corrente non equilibrata...
Page 42: ...Bedienungsanleitung ILCM 03A AC Leckstrompr fger t FR EN IT DE ES...
Page 48: ...5 3 1 2 Unwucht Leckstrom...
Page 52: ...9 halten Sie beim Einschalten des Ger ts die HOLD Taste gedr ckt...
Page 56: ...Manual de instrucciones ILCM 03A Pinza Amperim trica para Corrientes de Fuga FR EN IT DE ES...
Page 62: ...5 3 1 2 Corriente de Salida Fuera de Balance...
Page 69: ...12...
Page 70: ...13...