2. Connect the probes to the device to be measured.

3-3 Measuring DC Voltage (V)

1. Rotate the function selector to the 

V

 

'

 position.

2. Connect the probes to the device to be measured.

3-4 Measuring Continuity and Resistance (

 )

1. Rotate the function selector to the 

Ω□

 

position.

2. Verify that the power to the circuit under test is off. Connect test leads to the circuit to make the measurement.
3. The internal sounder will operate if the resistance of the circuit under test is less than 25

, it will then turn off if 

     the resistance is increased beyond 400

.

3-5 Measuring AC Current (A)

1. Rotate the function selector to the 

A

%

 position.

2. Refer to figure 2 and disconnect probes from probe test points.
3. Place the conductor anywhere within the sensor zone as shown in the shaded areas of Fig. 2.

Summary of Contents for ICM A7

Page 1: ...ISO TECH ICM A7 ELECTRICAL TESTER INSTRUCTION MANUAL...

Page 2: ......

Page 3: ...y and prevent damage to this product or any products Connected to it To avoid potential hazards use the product only as specified CAUTION These statements identify conditions or practices that could r...

Page 4: ...a voltage source probe tip sleeves are supplied 3 To avoid electric shock do not operate this product in high temperature or high humidity environment 4 Do not use the tester if it or its test leads...

Page 5: ...igit display maximum reading of 1999 with autopolarity and decimal point indicators 2 Measurement Function Dial Rotate the function dial to select V V and A function 3 Display Hold To hold the reading...

Page 6: ...Figure 1 4 1 3 2 5...

Page 7: ...play OL is displayed for A and functions shows the real value for V function Low voltage indicator is displayed Replace the battery when the indicator appears in the display Automatic power off time A...

Page 8: ...th battery removed Temperature Coefficient 0 2 x Specified accuracy 18 28 Operating Altitude 2000m 6562ft Battery Type and Life ALKALINE PP3 9V 250 hours Overvoltage Category IEC 1010 600V CAT Polluti...

Page 9: ...Voltage Function Range Resolution Accuracy Overload Protection V 600V rms 1V 1 5 3 digits 40Hz 500Hz 600V rms V 600V 1V 1 0 2 digits 600V rms Input impedance 1M less than 100pF 2 Resistance Continuit...

Page 10: ...e of the circuit under test is less than 25 It will then turn off if the resistance is increased beyond 400 3 AC Current Function Range Resolution Accuracy Overload Protection A 200A 0 1A 3 0 3 digits...

Page 11: ...ecome unstable and indicate large errors 2 Make sure that the battery is properly connected 3 The instrument should only be operated between 0 50 and at less than 80 R H 4 Do not use or store this ins...

Page 12: ...r to the circuit under test is off Connect test leads to the circuit to make the measurement 3 The internal sounder will operate if the resistance of the circuit under test is less than 25 it will the...

Page 13: ...OK OK OK OK Figure 2...

Page 14: ...her cleaning use a soft cloth or paper towel dampened with water You can use a 75 isopropyl alcohol solution for more efficient cleaning CAUTION To avoid damage to the surface of the meter do not use...

Page 15: ...Figure 3...

Page 16: ......

Reviews: