SfAR-S-6TI User Manual
www.ismacontrolli.com
DMP272en | 3rd Issue rev. 2 | 05/2022
page 7 of 26
T thermocouple operation range
-200°C to +400°C
N thermocouple operation range
-200°C to +1300°C
S thermocouple operation range
-0°C to +1700°C
R thermocouple operation range
-0°C to +1700°C
B thermocouple operation range
-0°C to +1800°C
Resistance
From 0 to 8000 Ω (1 Ω resolution)
Voltage
0 to 2048 mV (0.1 mV resolution)
0 to 256 mV (0.01 mV resolution)
Voltage measurement error
±0.05 % of the measurement range
Measuring current
~250 μA
Measurement resolution
0.1°C
Measurement error
±0.5°C
Cold junction temperature
measurement
-55°C to +100°C
Measurement error of the cold junction
temperature
±1°C
DAC processing time
150 ms/channel
Isolation
1000 V DC
Digital Outputs
Maximum current and voltage
500 mA/55 V
Isolation
1000 V DC
Temperature
Work
-10°C to +50°C (14°F to 122°F)
Storage
-40°C to +85°C (-40°F to 185°F)
Connectors
Power supply
2 pin
Communication
3 pin
Inputs and outputs
2 x 10 pin
Quick connector
IDC10
Configuration
mini USB
Size
Height
119.1 mm (4.689 in)
Length
110.9 mm (4.366 in)