![iSi ISI-799 Service Manual Download Page 5](http://html1.mh-extra.com/html/isi/isi-799/isi-799_service-manual_2092808005.webp)
5.
LB/KG (LB and KG Switch)
Enable or disable LB and KG switching.
6.
ZRTRBN (Zero tracking band)
OFF, 0.5d, 1d, or 3d can be entered.
7.
ZRANGE (Zero range)
2% or 100% can be entered.
8.
AUZERO (Auto zero)
If ZERO is selected means automatically zero the indicator when power on. If
NOZERO is selected means return to the zero position of ZERO CALIBRATION.
9.
MOTBAN (Motion band)
If motion value is within motion band, STABLE lights.
10.
FILT (Digital filter)
The bigger number, the stronger filter.
11.
CKWCOD(Check weight condition)
0:
Check on Counting mode and unstable.
1:
Check on Counting mode and stable.
2:
Check on Weighing mode and unstable.
3:
Check on Weighing mode and stable.
4:
Disable check function.
12.
PCS(Counting mode)
Enable Counting mode or not.
1.3.4 Calibration menu
CONFIG CALIB SERIAL PFORMT ALGOUT DIG IO
TEST VE
RS
ZERO
SPAN
CAL
Number
CAL
Return to weigh mode
Parameter Descriptions:
1.
ZERO (Zero Calibration)
Zero calibrates A/D count value.
2.
SPAN (Span Calibration)
Edit the test weigh value, then press “ENTER” key.