Since the sampler may encounter more than one error during a pro-
gram, it does not display the messages continuously. Instead, it
records each error and the time it occurred in the Sampling Results
and Combined Results report. It alerts you to the recorded errors by
displaying the message, “ERRORS HAVE OCCURRED DURING
PROGRAM.” (See
Table
for a description of the errors and events.)
Fatal Error
Screen
will only show up if you are using a multiple bottle config-
uration and the distributor system fails. Failing the test can indi-
cate any of the following conditions:
• The distributor arm is not attached to the sampler.
• The stop arm on the distributor shaft extension is missing or
broken.
• The stop inside the refrigerator is missing or broken.
• The distributor drive mechanism inside the controller is mal-
functioning.
Warning Messages
Occasionally, the sampler displays a warning message to alert you
to a potential problem. There are two warnings. The first warning,
“REPLACE PUMP TUBE,” reminds you to check the condition of
the pump tube. For more information, see Replacing the Pump
Tube on page 70. The second warning reminds you that the inter-
nal battery is near expiration and should be replaced. You can find
instructions for replacing the battery in Replacing the Internal
Battery on page 75.
Interrupting a Running Program
You can interrupt a sampling program by pressing the
Stop
key
while the sampler is waiting for the next sample event. Pressing
Stop
places the sampler into Manual Paused operation and records
a “MANUAL PAUSE” in the sample event log.
In this state, the sampler allows you to access several functions.
You can modify the program, take a grab sample, etc., with out ad-
versely affecting the running program. Modifying the program is
limited to adjusting the pacing intervals, enable conditions, sample
volume, and suction head. The basic program structure, such as the
type of pacing, can not be changed.
While in the manual paused state, the sampling program continues
to operate as normal, with the exception of taking samples. If a
sample was to be taken, it is skipped. The sampler records “SAM-
PLE SKIPPED” in the sample event log and continues to operate
as normal.
The manual paused state displays a scrolling menu with up to
twelve options. Use the Arrow keys to scroll through the manual
paused options and the Enter key to make a selection. Return to
the running program by selecting “RESUME PROGRAM” or press-
ing the
Stop
key. If you do not make a selection, the sampler auto-
matically resumes the original sampling program five minutes
after the last key was pressed.
Warning Messages
Program Done Messages
At the completion of a multi-bottle
sampling program, the controller
“parks” the distributor towards
the back of the refrigerator. The
parked position allows for easier
access to the bottle rack.
l
WARNING: REPLACE
PUMP TUBE
m
WARNING!
REPLACE INTERNAL
BATTERY BEFORE
1-OCT-97
n
PROGRAM
IS DONE
Standard program
completed.
o
PROGRAM: “EXTENDED 1”
IS DONE
Extended program
completed.
Note
Summary of Contents for 6700SR
Page 2: ......
Page 4: ......
Page 18: ...6700SR Refrigerated Sampler iv...
Page 148: ...6700SR Refrigerated Sampler 130 Appendix B Material Safety Data Sheets...
Page 149: ...6700SR Refrigerated Sampler 131...
Page 150: ...6700SR Refrigerated Sampler 132...
Page 151: ...6700SR Refrigerated Sampler 133...
Page 152: ...6700SR Refrigerated Sampler 134...
Page 153: ...6700SR Refrigerated Sampler 135...
Page 154: ...6700SR Refrigerated Sampler 136...
Page 155: ...6700SR Refrigerated Sampler 137...
Page 156: ...6700SR Refrigerated Sampler 138...
Page 157: ...6700SR Refrigerated Sampler 139...
Page 158: ...6700SR Refrigerated Sampler 140...
Page 159: ...6700SR Refrigerated Sampler 141...
Page 160: ...6700SR Refrigerated Sampler 142...
Page 161: ...6700SR Refrigerated Sampler 143...
Page 162: ...6700SR Refrigerated Sampler 144...
Page 163: ...6700SR Refrigerated Sampler 145...
Page 164: ...6700SR Refrigerated Sampler 146...
Page 165: ...6700SR Refrigerated Sampler 147...
Page 166: ...6700SR Refrigerated Sampler 148...
Page 167: ...6700SR Refrigerated Sampler 149...
Page 168: ...6700SR Refrigerated Sampler 150...
Page 174: ...6700SR Refrigerated Sampler 156 Appendix D Replacement Parts REFRIGERATOR ASSEMBLY FRONT BACK...
Page 175: ...6700SR Refrigerated Sampler 157 REFRIGERATOR ASSEMBLY...
Page 176: ...6700SR Refrigerated Sampler 158 MODEL 6700 CONTROLLER INSIDE...
Page 177: ...6700SR Refrigerated Sampler 159 MODEL 6700 CONTROLLER INSIDE...
Page 178: ...6700SR Refrigerated Sampler 160 MODEL 6700 CONTROLLER OUTSIDE...
Page 179: ...6700SR Refrigerated Sampler 161 MODEL 6700 CONTROLLER OUTSIDE...
Page 180: ...6700SR Refrigerated Sampler 162 MODEL 6700 PUMP GEAR CASE ASSEMBLY...
Page 188: ...6700SR Refrigerated Sampler 170...
Page 190: ......