6700SR Refrigerated Sampler
50
Samplers with an attached YSI Sonde will display the parameter
readings selected during the YSI hardware setup. The YSI screens
alternate with the sampler and module screens.
The Module and YSI readings are updated:
• Every 5 seconds when not running a program and not display-
ing the reading.
• Every 5 seconds when running a program and the backlight is
on.
• Every 30 seconds when running a program and the reading is
used for pacing or enabling the sampler.
• At the programmed storage interval when running a program
and the reading is not used for pacing or enabling the sampler.
• Every second when displaying the reading during program-
ming.
An
❋
(asterisk) appears next to the reading if the module or YSI
Sonde was unable to take a reading. If an asterisk appears, the
reading displayed is the last available reading.
If an entire reading is filled with asterisks, the value exceeds the
number of characters that can be displayed. The sampler stores the
actual reading in memory, so it may be collected later.
For those programs that have delayed or scheduled start times, the
readings will be displayed while waiting for the start time. Keep in
mind that any totalizer values will be reset at the start time. Mod-
ule and YSI readings are also displayed after a program is com-
plete, until a key is pressed.
It is not necessary to run a program to view module or YSI read-
ings. Real time data can be viewed on the display by selecting
“VIEW REPORT” from the main menu. See Viewing the Data on
page 54.
Error Messages
If the sampler detects an error that prevents it from taking a sam-
ple or continuing the program, it displays an error message.
There are messages for the following errors:
• Pump Jammed
• No Liquid Detected
• Distributor Jammed
• No More Liquid
• Probable Overflow
• Pump Latch open
• Power Failed
• Sampler Shut Off
• User Stopped
• Bottle Full
• No Distributor Arm
• Sample In Progress
BOTTLE 2
IN 00:14:39
150 gpm
0.82 ft
00001438200 gal
Y-pH:
8.1
Y-D.O:
4.90
mg/l
Y-TDS:
1.000
g/l
Y-TEMP:
29.2
C
Error Messages
i
BOTTLE 2
IN 00:12:00
❋❋❋❋❋
gpm
5.73 ft
00001457600 gal
j
ERRORS HAVE OCCURED
DURING PROGRAM
k
FATAL ERROR
NO DISTRIBUTOR ARM
Summary of Contents for 6700SR
Page 2: ......
Page 4: ......
Page 18: ...6700SR Refrigerated Sampler iv...
Page 148: ...6700SR Refrigerated Sampler 130 Appendix B Material Safety Data Sheets...
Page 149: ...6700SR Refrigerated Sampler 131...
Page 150: ...6700SR Refrigerated Sampler 132...
Page 151: ...6700SR Refrigerated Sampler 133...
Page 152: ...6700SR Refrigerated Sampler 134...
Page 153: ...6700SR Refrigerated Sampler 135...
Page 154: ...6700SR Refrigerated Sampler 136...
Page 155: ...6700SR Refrigerated Sampler 137...
Page 156: ...6700SR Refrigerated Sampler 138...
Page 157: ...6700SR Refrigerated Sampler 139...
Page 158: ...6700SR Refrigerated Sampler 140...
Page 159: ...6700SR Refrigerated Sampler 141...
Page 160: ...6700SR Refrigerated Sampler 142...
Page 161: ...6700SR Refrigerated Sampler 143...
Page 162: ...6700SR Refrigerated Sampler 144...
Page 163: ...6700SR Refrigerated Sampler 145...
Page 164: ...6700SR Refrigerated Sampler 146...
Page 165: ...6700SR Refrigerated Sampler 147...
Page 166: ...6700SR Refrigerated Sampler 148...
Page 167: ...6700SR Refrigerated Sampler 149...
Page 168: ...6700SR Refrigerated Sampler 150...
Page 174: ...6700SR Refrigerated Sampler 156 Appendix D Replacement Parts REFRIGERATOR ASSEMBLY FRONT BACK...
Page 175: ...6700SR Refrigerated Sampler 157 REFRIGERATOR ASSEMBLY...
Page 176: ...6700SR Refrigerated Sampler 158 MODEL 6700 CONTROLLER INSIDE...
Page 177: ...6700SR Refrigerated Sampler 159 MODEL 6700 CONTROLLER INSIDE...
Page 178: ...6700SR Refrigerated Sampler 160 MODEL 6700 CONTROLLER OUTSIDE...
Page 179: ...6700SR Refrigerated Sampler 161 MODEL 6700 CONTROLLER OUTSIDE...
Page 180: ...6700SR Refrigerated Sampler 162 MODEL 6700 PUMP GEAR CASE ASSEMBLY...
Page 188: ...6700SR Refrigerated Sampler 170...
Page 190: ......