6700SR Refrigerated Sampler
75
Replacing the Internal Battery
A Lithium battery housed inside the controller maintains power to
the sampler’s memory when the controller is disconnected from a
power source. If it discharges completely, the sampler win lose all
program settings and all data stored in memory when the external
power source is disconnected. The battery requires replacement ev-
ery five years. The sampler displays a warning within one month
of its expected life to alert you to the battery’s impending expira-
tion. The warning appears each time you turn the sampler on until
you reset the alarm. When you reset the alarm, the sampler resets
the warning to a date 59 months from the current date.
To replace the internal battery:
1. Open the controller case (see Opening the Controller Case on
2. Lift the desiccant box from the case (
Figure
3. Remove the EMI shield and grounding straps by loosening
the five
1
/
4
-inch stop nuts.
4. Disconnect the module, liquid detector, and pump count con-
nectors from the main circuit board (
Figure
).
5. Remove the three screws and the
1
/
4
-inch stand-off attaching
the main circuit board to the case.
6. Detach the main circuit board from the keypad connector by
pulling gently upward on the edges of the board next to the
connector. Pull the main board up until the keypad connector
is almost even with the edge of the case. Next, detach the main
circuit board from the connector board.
7. Desolder the lithium battery, using care to prevent damage
to the circuit board. Insert the replacement battery and sol-
der it in place.
8. Reconnect the main circuit board to the connector board be-
fore reconnecting it to the keypad. Replace the
1
/
4
-inch stand-
off and the three screws, and reconnect the module, liquid de-
tector, and pump count connectors.
9. Replace the EMI shield and grounding straps. Reconnect the
two distributor connectors.
10.Ensure that all connections and hardware are secure. Re-
place the desiccant and close the case.
11.Reset the battery alarm.
Battery Warning
Isco recommends you replace
the battery every five years, as
soon as the battery warning
appears. To insure that the re-
placement meets Isco’s speci-
fications, use only Isco re-
placement batteries. (See Ap-
pendix D, Replacement Parts.)
WARNING!
REPLACE INTERNAL
BATTERY BEFORE
17-OCT-97
Note
The circuit boards can be dam-
aged from a discharge of static
electricity. Always use an anti-
static mat when opening the con-
troller to protect the circuit
boards.
Poor soldering techniques can
easily damage a printed circuit
board. The battery should be re-
placed by a skilled technician.
CAUTION
Note
Summary of Contents for 6700SR
Page 2: ......
Page 4: ......
Page 18: ...6700SR Refrigerated Sampler iv...
Page 148: ...6700SR Refrigerated Sampler 130 Appendix B Material Safety Data Sheets...
Page 149: ...6700SR Refrigerated Sampler 131...
Page 150: ...6700SR Refrigerated Sampler 132...
Page 151: ...6700SR Refrigerated Sampler 133...
Page 152: ...6700SR Refrigerated Sampler 134...
Page 153: ...6700SR Refrigerated Sampler 135...
Page 154: ...6700SR Refrigerated Sampler 136...
Page 155: ...6700SR Refrigerated Sampler 137...
Page 156: ...6700SR Refrigerated Sampler 138...
Page 157: ...6700SR Refrigerated Sampler 139...
Page 158: ...6700SR Refrigerated Sampler 140...
Page 159: ...6700SR Refrigerated Sampler 141...
Page 160: ...6700SR Refrigerated Sampler 142...
Page 161: ...6700SR Refrigerated Sampler 143...
Page 162: ...6700SR Refrigerated Sampler 144...
Page 163: ...6700SR Refrigerated Sampler 145...
Page 164: ...6700SR Refrigerated Sampler 146...
Page 165: ...6700SR Refrigerated Sampler 147...
Page 166: ...6700SR Refrigerated Sampler 148...
Page 167: ...6700SR Refrigerated Sampler 149...
Page 168: ...6700SR Refrigerated Sampler 150...
Page 174: ...6700SR Refrigerated Sampler 156 Appendix D Replacement Parts REFRIGERATOR ASSEMBLY FRONT BACK...
Page 175: ...6700SR Refrigerated Sampler 157 REFRIGERATOR ASSEMBLY...
Page 176: ...6700SR Refrigerated Sampler 158 MODEL 6700 CONTROLLER INSIDE...
Page 177: ...6700SR Refrigerated Sampler 159 MODEL 6700 CONTROLLER INSIDE...
Page 178: ...6700SR Refrigerated Sampler 160 MODEL 6700 CONTROLLER OUTSIDE...
Page 179: ...6700SR Refrigerated Sampler 161 MODEL 6700 CONTROLLER OUTSIDE...
Page 180: ...6700SR Refrigerated Sampler 162 MODEL 6700 PUMP GEAR CASE ASSEMBLY...
Page 188: ...6700SR Refrigerated Sampler 170...
Page 190: ......